Application of an Amplitude-integrated EEG Monitor Cerebral Function Monitor to Neonates
Published on: September 6, 2017
Comparing pulse rate measurement in newborns using conventional and dry-electrode ECG monitors
Eris van Twist1, Hylke H Salverda2, Arjan B Te Pas2
1Pediatric Intensive Care Unit, Department of Pediatrics and Pediatric Surgery, Erasmus MC, Sophia Children's Hospital, University Medical Center Rotterdam, Rotterdam, the Netherlands.
Aim:
Heart rate (HR) is the most important parameter to evaluate newborns' clinical condition and to guide intervention during resuscitation at birth. The present study aims to compare the accuracy of NeoBeat dry-electrode ECG for HR measurement with conventional ECG and pulse oximetry (PO).
Methods:
Newborns with a gestational age ≥32 weeks and/or birth weight ≥1.5 kg were included when HR evaluation was needed. HR was simultaneously measured for 10 min with NeoBeat, PO and conventional ECG.
Results:
A total of 18 infants were included (median (IQR) gestational age 39 (36-39) weeks and birth weight 3 150 (2 288-3 859) grams). Mean (SD) duration until NeoBeat obtained a reliable signal was 2.5 (9.0) s versus 58.5 (171.0) s for PO. Mean difference between NeoBeat and ECG was 1.74 bpm (LoA -4.987-8.459 and correlation coefficient 0.98). Paired HR measurements over 30-s intervals revealed no significant difference between NeoBeat and ECG. The positive predictive value of a detected HR <100 bpm by NeoBeat compared with ECG was 54.84%, negative predictive value 99.99%, sensitivity 94.44%, specificity 99.99% and accuracy 99.85%.
Conclusions:
HR measurement with NeoBeat dry-electrode ECG at birth is reliable and accurate.
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Special considerations while measuring pulse
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...
ECG Interpretation of Rhythms
Components of the Electrocardiogram
The primary components of a normal ECG waveform in Normal sinus rhythm(NSR) include the P wave, PR interval, QRS complex, ST segment, T wave, and occasionally a U wave.
ECG waveforms are divided by vertical and horizontal lines at standard intervals.
The horizontal axis measures time and rate, and the vertical axis measures amplitude or voltage....