Graft Resistance Difference after HVAD to HeartMate 3 Left Ventricular Assist Device Exchange
Kevin Bourque1, Matthew J Martin1, Daniel I Harjes1
1Heart Failure Division, Abbott, Burlington, Massachusetts.
The Annals of Thoracic Surgery
|January 6, 2022
Summary
Replacing the HeartWare Ventricular Assist Device (HVAD) with the HeartMate 3 may require higher pump speeds if the old HVAD outflow graft is retained. Complete conversion to the HeartMate 3 graft is preferred for optimal performance.
Area of Science:
- Cardiovascular Engineering
- Medical Device Technology
Background:
- The discontinuation of the HeartWare Ventricular Assist Device (HVAD) necessitates increased use of the HeartMate 3 for left ventricular assist device replacements.
- Retaining parts of the HVAD's 10-mm outflow graft during HeartMate 3 implantation may require higher device operating speeds compared to using the standard 14-mm HeartMate 3 graft.
Purpose of the Study:
- To evaluate the hemodynamic impact of using a remnant HVAD outflow graft with the HeartMate 3 left ventricular assist device.
- To determine the effects of partial HVAD graft retention on HeartMate 3 performance parameters, including rotor speed and power consumption.
Main Methods:
- An in vitro study was conducted using steady-state flow conditions.
- Samples included HeartWare Ventricular Assist Device (HVAD), HeartMate 3, and hybrid HVAD/HeartMate 3 grafts of varying lengths (125-375 mm).
- Flow rates from 3.0 to 6.0 L/min were applied to simulate clinical scenarios.
Main Results:
- At clinically relevant flow rates (4.3 and 6.0 L/min), HeartMate 3 rotor speeds increased from 5350 to 5700 rpm and 6350 to 6900 rpm, respectively, when using the full HVAD graft compared to the HeartMate 3 graft.
- Corresponding power consumption increased by 16% (3.7 to 4.3 W) and 24% (5.5 to 6.8 W) at these flow rates.
Conclusions:
- While no severe immediate consequences were identified, using a remnant HVAD graft necessitates higher HeartMate 3 rotor speeds.
- This can lead to reduced patient battery runtime and increased power consumption.
- Complete replacement with the HeartMate 3 graft is the preferred approach from an engineering perspective for optimal device function and patient benefit.
Related Concept Videos
Cardiac Catheterization III: Left Heart Catheterization
256
Left heart catheterization is an invasive diagnostic procedure used to evaluate the function and structure of the left side of the heart. It is generally performed to diagnose and treat cardiovascular conditions such as valve abnormalities, coronary artery disease, and congenital heart defects.Diagnostic and therapeutic purposesLeft heart catheterization serves various diagnostic and therapeutic purposes, including:Assessing coronary artery bypass grafts.Evaluating coronary artery disease in...
256
Cardiomyopathy V: Interprofessional Care
60
Managing cardiomyopathy involves addressing underlying or precipitating causes, treating heart failure with medications, and implementing dietary changes and a balanced exercise and rest regimen.Lifestyle ModificationsCardiomyopathy patients should adopt a low-sodium diet to reduce fluid retention and manage heart failure. A personalized exercise and rest plan helps maintain physical fitness without overstraining the heart. Avoiding alcohol and tobacco is essential to prevent further damage to...
60
Cardiac Catheterization II: Right Heart Catheterization
229
Right Heart Catheterization: An OverviewRight heart catheterization is an invasive diagnostic procedure that measures right-sided cardiac and pulmonary artery pressures, calculates cardiac output, and identifies intracardiac shunts. It provides detailed hemodynamic data essential for diagnosing and managing various cardiovascular conditions, such as pulmonary hypertension.Access SitesCommon access sites for right heart catheterization include the internal jugular vein in the neck region, the...
229