MaxiCalc: A tool for online dosimetric evaluation of source-tracking based treatment verification in HDR
Maximilian D Hanlon1, Ryan L Smith1, Rick D Franich2
1School of Science, RMIT University, Melbourne, Australia; Alfred Health Radiation Oncology, The Alfred, Melbourne, Australia.
Purpose:
Source tracking is becoming a more widely used approach in HDR brachytherapy treatment verification. While it provides a sensitive method to detect deviations from the treatment plan during delivery, it does not show the clinical significance of any detected changes. By incorporating a tool that calculates volumetric doses and DVH indices from measurements, source tracking systems can be expanded to assess dosimetric significance of any deviations from the plan.
Methods:
The source tracking dose calculation tool, MaxiCalc, was developed in MATLAB. Validation was performed by comparing doses and DVH indices calculated in MaxiCalc to those calculated by the clinical TPS, for several test plans and 10 clinical plans. Clinical implementation was demonstrated by calculating volumetric doses from a clinical source tracking event.
Results:
MaxiCalc showed excellent agreement with the clinical TPS for point and volumetric doses (mean difference < 0.01% and 0.1% respectively). MaxiCalc calculates dosimetrically equivalent plans to the TPS with agreement < 0.3% for all DVH indices except PTV V200%. Small differences seen for the clinical source tracking event were consistent with the known tracking uncertainties enabling them to be quantified for clinical decision making. Calculations are fast, enabling real-time use.
Conclusions:
MaxiCalc is an independent tool that calculates doses and DVH indices from dwells measured with any clinical HDR brachytherapy source tracking system. This extends the capabilities of source tracking systems from determining discrepancies in positions or times during delivery to assessing the dosimetric impact of any detected deviations, allowing for more comprehensive treatment verification and evaluation.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
10:33Expedited Radiation Biodosimetry by Automated Dicentric Chromosome Identification ADCI and Dose Estimation
Published on: September 4, 2017
Related Concept Videos
Radiation: Applications
The average...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...