Development and Implementation of One-Step, Broad-Spectrum, High-Sensitivity Drug Screening by Tandem Mass
Merih T Tesfazghi1, Rick Bardelmeier2, Al N Saunders2
1Department of Pathology, Rush University Medical Center, Chicago, IL.
A new liquid chromatography-tandem mass spectrometry (LC-MS/MS) drug screen improves accuracy in pediatric urine samples. This method detects more compounds than traditional immunoassays, aiding clinical decisions.
Area of Science:
- Clinical Chemistry
- Forensic Toxicology
- Pediatric Drug Monitoring
Background:
- Immunoassay drug screening is prevalent in pediatric settings.
- Common issues include false positives/negatives due to antibody cross-reactivity and dilute urine.
- Accurate drug exposure data is critical for pediatric clinical and medico-legal contexts.
Purpose of the Study:
- To develop and validate a liquid chromatography-tandem mass spectrometry (LC-MS/MS) assay for pediatric urine drug screening.
- To compare the performance of LC-MS/MS with traditional immunoassay methods.
- To analyze the frequency and types of drug detection in a pediatric population post-implementation.
Main Methods:
- Development and characterization of a broadly targeted LC-MS/MS assay for 38 compounds.
- Retrospective analysis of 3985 pediatric urine drug screens (1663 pre-implementation, 2322 post-implementation).
- Comparison of immunoassay results with LC-MS/MS confirmation and post-implementation LC-MS/MS data.
Main Results:
- Immunoassays showed high rates of presumptive positives (23% general pediatric, 37% nursery), with significant discrepancies upon mass spectrometry confirmation (85% general pediatric, 40% nursery confirmed).
- Post-LC-MS/MS implementation, 31% of general pediatric and 18% of nursery samples tested positive.
- LC-MS/MS identified a broader range of compounds, notably under-detecting THC and over-detecting cocaine with immunoassays in the nursery population.
Conclusions:
- A broadly targeted, sensitive LC-MS/MS assay offers superior detection of multiple drug compounds in a single step compared to immunoassay screen-then-confirm.
- Accurate and rapid results from LC-MS/MS facilitate timely patient management in diverse clinical settings.
- LC-MS/MS provides a more comprehensive and accurate assessment of pediatric drug exposure.