Background:
Impedance pneumography (IP) is the current device-driven method used to measure respiratory rate (RR) in hospitalized patients. However, RR alarms are common and contribute to alarm fatigue. While RR derived from electrocardiographic (ECG) waveforms hold promise, they have not been compared to the IP method.
Purpose:
Study examined the agreement between the IP and combined-ECG derived (EDR) for normal RR (≥12 or ≤20 breaths/minute [bpm]); low RR (≤5 bpm); and high RR (≥30 bpm).
Methodology:
One-hundred intensive care unit patients were included by RR group: (1) normal RR (n = 50; 25 low RR and 25 high RR); (2) low RR (n = 50); and (3) high RR (n = 50). Bland-Altman analysis was used to evaluate agreement.
Results:
For normal RR, a significant bias difference of -1.00 + 2.11 (95% CI -1.60 to -0.40) and 95% limit of agreement (LOA) of -5.13 to 3.13 was found. For low RR, a significant bias difference of -16.54 + 6.02 (95% CI: -18.25 to -14.83) and a 95% LOA of -28.33 to - 4.75 was found. For high RR, a significant bias difference of 17.94 + 12.01 (95% CI: 14.53 to 21.35) and 95% LOA of -5.60 to 41.48 was found.
Conclusion:
Combined-EDR method had good agreement with the IP method for normal RR. However, for the low RR, combined-EDR was consistently higher than the IP method and almost always lower for the high RR, which could reduce the number of RR alarms. However, replication in a larger sample including confirmation with visual assessment is warranted.
Related Concept Videos
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement:
Physical Assessment of the Respiratory Tract II: Inspection
Chest Configuration
The chest configuration...
Respiratory Volumes and Capacities I
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...