Organ dose in cardiac dual-energy computed tomography: a Monte Carlo study
Delaram Pakravan1, Farshid Babapour Mofrad2, Mohammad Reza Deevband3
1Department of Medical Radiation Engineering, Science and Research Branch, Islamic Azad University, Tehran, Iran.
Abstract:
Dual-energy computed tomography (DECT) has appeared as a novel approach with the aim of evaluating artery-related diseases. With the advent of DECT, concerns have been raised about the induction of diseases such as cancer due to high radiation exposure of patients. Therefore, the dose received by patients in DECT should be considered. The parameter most commonly used for patient dosimetry is the effective dose (ED). The purpose of this study is to model and validate a DECT scanner by a developed MCNP Monte Carlo code and to calculate the organ doses, the ED, and the conversion factor (k-factor) used in determining ED in the cardiac imaging protocol. To validate the DECT scanner simulation, a standard dosimetry body phantom was modeled in two radiation modes of single energy CT and DECT. The results of simulated CT dose index (CTDI) were compared with those of ImPACT or measurement data. Then dosimetry phantom was replaced by the male and female ORNL phantoms and the organ doses were calculated. The organ doses were also calculated by ImPACT dose software. In the initial validation stage, the minimum and maximum observed relative differences between results of MNCP simulation and measured were 2.77% and 5.79% for the central CTDI and 1.91% and 5.83% for the averaged peripheral CTDI, respectively. The mean ED of simulation and the ImPACT were 3.23 and 5.55 mSv/100 mAs, and the mean k-factor was 0.016 and 0.032 mSv mGy-1 cm-1 in the male and female phantoms, respectively. The k-factor obtained for males is close to the currently used k-factor, but the k-factor for females is almost twice.
More Related Videos
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Imaging Studies III: Computed Tomography
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram