Dual k-space and image-space post-processing for field-cycling MRI under low magnetic field stability and homogeneity
Gonzalo G Rodriguez1, Alejandro Salvatori1, Esteban Anoardo1
1Laboratorio de Relaxometría y Técnicas Especiales (LaRTE), Grupo de Resonancia Magnética Nuclear, FaMAF - Universidad Nacional de Córdoba & IFEG-CONICET, Córdoba, Argentina.
Abstract:
We present a method to correct artifacts typically present in images acquired in field-cycled MRI experiments under poor magnetic field spatial-homogeneity and time-stability conditions. The proposed method was tested in both simulated and experimental data. The experiments were performed using a fast field-cycling MRI relaxometer of own design, based on a current-driven variable-geometry electromagnet. Current instability-induced artifacts in the images were mitigated through a phase correction array resulted from entropy and background minimization. Image distortions due to magnetic field inhomogeneity were compensated through two different approaches, involving a previous determination of the magnetic field homogeneity-map, or an experimental protocol where two images are acquired with inverted readout gradient polarity. Results show that images acquired at extreme conditions can be successfully improved, thus strengthening the possibilities for both low-cost MRI devices and faster field-switched MRI systems.
Related Concept Videos
Magnetic Resonance Imaging
NMR Spectrometers: Resolution and Error Correction
Atomic Nuclei: Nuclear Relaxation Processes
Double Resonance Techniques: Overview
Spin decoupling is usually achieved by...