Related Experiment Video
Updated: Oct 6, 2025

Simultaneous Data Collection of fMRI and fNIRS Measurements Using a Whole-Head Optode Array and Short-Distance Channels
Published on: October 20, 2023
Comparison of forehead and finger oximetry sensors during the six minute walk test
Lucy Robertson1, Rachel Lowry1, Karl Sylvester1
1RinggoldID:2144Royal Papworth Hospital, Cambridge, UK.
Background:
Measurement of oxygen saturation (SpO2) during the 6 minute walk test (6MWT) could be impacted by the measurement site.
Aims:
To compare SpO2 and heart rate (HR) between forehead and finger sensors during the 6MWT. Sensor readings were also to be compared for signal quality and with capillary blood gas (CBG) pre and post 6MWT.
Method:
80 subjects with pulmonary vascular disease (PVD) and/or interstitial lung disease (ILD) performed the 6MWT. Pulse oximetry was recorded at 30 s intervals. CBG was taken pre and post 6MWT to determine capillary oxygen saturation (SCO2).
Results:
The forehead sensor recorded higher values for SpO2 (p < 0.001) and HR (p < 0.01) compared with the finger sensor during the 6MWT. For both sensors, the demonstrated bias compared to CBG post 6MWT was higher and more variable in subjects who desaturated. During the 6MWT there was a higher occurrence (p < 0.001) of poor signal quality in the finger sensor compared with the forehead sensor.
Conclusion:
This study suggests that the sensor site can impact pulse oximetry readings. The variance in bias suggests pulse oximetry may not accurately reflect SCO2 measurements particularly in subjects who desaturate during 6MWT.
More Related Videos
04:20Integration of Brain Tissue Saturation Monitoring in Cardiopulmonary Exercise Testing in Patients with Heart Failure
Published on: October 1, 2019
09:33Using Near-Infrared Spectroscopy Wearable Devices to Identify Central Versus Peripheral Limitations During Exercise
Published on: December 19, 2024
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Assessment of blood pressure in brachial artery(two-step method)
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure: