Performing Intracochlear Electrocochleography During Cochlear Implantation
Published on: March 8, 2022
Increasing the reliability of real-time electrocochleography during cochlear implantation: a standardized guideline
K Schuerch1,2, M Waser2, G Mantokoudis2
1Hearing Research Laboratory, ARTORG Center for Biomedical Engineering Research, University of Bern, Bern, Switzerland.
Purpose:
Electrocochleography (ECochG) measures electrical potentials generated by the inner ear in response to acoustic stimulation. Real-time (rt) recordings are increasingly used during cochlear implant (CI) surgeries to monitor the inner ear function. However, the performance of rt-ECochG is a delicate measurement procedure involving several pitfalls, which lead to inaccurate or invalid signal recordings in up to 20%. In order to use the technique routinely in CI candidates, an improvement in measurement reliability must be achieved.
Methods:
In our prospective study, we systematically investigated potential pitfalls and error sources during rt-ECochG recordings. We performed experiments (i) on a head and torso simulator, (ii) on a whole-head cadaver specimen, (iii) as well as in vivo during rt-ECochG recordings in CI recipients. After analyzing experiments i-iii, a standardized measurement procedure was developed. We followed this guideline in 10 CI recipients to test the measurement reliability.
Results:
Besides improper installation, surgical and patient-specific factors influenced the measured signal. In particular, the unattenuated presentation of the acoustic stimulus was of importance. We summarized our findings in a standardized guideline. Following this guideline, we measured successful intraoperative ECochG recordings in 9/10 patients.
Conclusions:
Our error analysis improved the understanding of successful rt-ECochG measurements. When following our proposed guideline, we achieved more reliable intraoperative ECochG recordings.