Study on the Reliability and Accuracy of Scolioscope, a New Digital Scoliometer
Georgios Krekoukias1,2,3, George A Koumantakis2, Vasileios S Nikolaou1
1Medical School, National and Kapodistrian University of Athens, 11527 Athens, Greece.
Abstract:
Early detection of scoliosis with school screening and quick, easy, and reliable assessment of its progress are of paramount importance in the management of patients. There have been several tools described, with the most common being the analog scoliometer. Most recently, smartphone applications have entered this area with and without the use of sleeves for the device. There is no research that has evaluated the accuracy of measurements both left and right in either digital or analog devices. In this study, we evaluated the reliability and validity of a new digital scoliometer called the Scolioscope. Thirty subjects were included for the intra-rater reliability study. ICC values >0.9 were calculated both for same-day and between-day measurements. The device was highly accurate with an average difference from the ones set on the sine bar of 0.03° for right-side measurements and 0.18° for the left. These measurements suggest a highly accurate and reliable tool.