Improvement of kilovoltage intrafraction monitoring accuracy through gantry angles selection
Loïc Vander Veken1, David Dechambre2, Steven Michiels1,3
1Institut de Recherche Experimentale et Clinique (IREC), Center of Molecular Imaging, Radiotherapy and Oncology (MIRO), Université Catholique de Louvain, 1200 Brussels, Belgium.
Abstract:
Kilovoltage intrafraction monitoring (KIM) is a method allowing to precisely infer the tumour trajectory based on cone beam computed tomography (CBCT) 2D-projections. However, its accuracy is deteriorated in the case of highly mobile tumours involving hysteresis. A first adaptation of KIM consisting of a prior amplitude based binning step has been developed in order to minimize the errors of the original model (phase-KIM). In this work, we propose enhanced methods (KIMsub-arc optimand phase-KIMsub-arc optim) to improve the accuracy of KIM and phase-KIM which relies on the selection of the optimal starting CBCT gantry angle. Aiming at demonstrating the interest of our approach, we carried out a simulation study and an experimental study: we compared the accuracy of the conventional versus sub-arc optim methods on simulated realistic tumour motions with amplitudes ranging from 5 to 30 mm in 1 mm increments. The same approach was performed using a lung dynamic phantom generating a 30 mm amplitude sinusoidal motion. The results show that for in-silico simulated motions of 10, 20 and 30 mm amplitude, the three-dimensional root mean square error (3D-RMSE) can be reduced by 0.67 mm, 0.91 mm, 0.94 mm and 0.18 mm, 0.25 mm, 0.28 mm using KIMsub-arc optimand phase-KIMsub-arc optimrespectively. Considering all in-silico simulated trajectories, the percentage of errors larger than 1 mm decreases from 21.9% down to 1.6% for KIM (p < 0.001) and from 6.6% down to 1.2% for phase-KIM (p < 0.001). Experimentally, the 3D-RMSE is lowered by 0.5732 mm for KIM and by 0.1 mm for phase-KIM. The percentage of errors larger than 1 mm falls from 39.7% down to 18.5% for KIM and from 23.2% down to 11.1% for phase-KIM. In conclusion, our method efficiently anticipates CBCT gantry angles associated with a significantly better accuracy by using KIM and phase-KIM.