Transience of interference in an immunoassay measuring serum levels of beta-D-glucan

Harry E Prince1, Hollis J Batterman1, Susan E Realegeno1

  • 1Quest Diagnostics, San Juan Capistrano, CA, USA.

Insights

Interference in 1-3-beta-D-glucan testing for invasive fungal infections is often transient. For most patients, a follow-up sample confirmed the initial interference was temporary, with a median resolution time of 8 days.

Area of Science:

  • Clinical chemistry
  • Mycology
  • Infectious diseases

Background:

  • 1-3-beta-D-glucan is a biomarker for invasive fungal infections.
  • Assays for this biomarker can be subject to interference, potentially impacting diagnostic accuracy.

Purpose of the Study:

  • To evaluate the transient nature of interference in 1-3-beta-D-glucan testing.
  • To determine the time course of interference resolution in patient samples.

Main Methods:

  • Analysis of results from 426 patients with an initial sample showing interference.
  • Comparison of initial interference samples with subsequent patient samples.
  • Assessment of interference status (positive/negative) in later samples.

Main Results:

  • Interference was observed in initial samples from 426 patients.
  • Follow-up samples indicated transient interference in 73% of these patients.
  • The median time between the initial interference sample and a resolved sample was 8 days.

Conclusions:

  • Interference in 1-3-beta-D-glucan testing is frequently transient.
  • This transient nature suggests that re-testing or careful interpretation of initial results may be warranted.
  • Understanding interference patterns aids in reliable diagnosis of invasive fungal infections.