DOAC Dipstick Testing Can Reliably Exclude the Presence of Clinically Relevant DOAC Concentrations in Circulation
Sandra Margetić1, Ivana Ćelap1, Arijana Lovrenčić Huzjan2
1Department of Clinical Chemistry, Sestre Milosrdnice University Hospital Center, Zagreb County, Croatia.
Qualitative urine testing with DOAC Dipstick can reliably exclude clinically relevant plasma levels of direct oral anticoagulants (DOACs) at specific thresholds. This aids in managing anticoagulant therapy by confirming the absence of significant drug concentrations.
Area of Science:
- Clinical Pharmacology
- Diagnostic Testing
- Pharmacokinetics
Background:
- Determining clinically relevant plasma levels of direct oral anticoagulants (DOACs) is crucial in specific clinical scenarios.
- Existing methods for DOAC quantification can be complex or time-consuming, necessitating simpler diagnostic approaches.
Purpose of the Study:
- To evaluate the utility of qualitative urine testing using DOAC Dipstick to exclude significant plasma concentrations of DOACs (≥30 ng/mL).
- To determine the negative predictive value (NPV) and other diagnostic parameters of the DOAC Dipstick for direct factor Xa inhibitors (DXIs) and direct thrombin inhibitors (DTIs).
Main Methods:
- A prospective single-center cohort study included 128 patients treated with apixaban, rivaroxaban, or dabigatran.
- DOAC urine testing was performed using DOAC Dipstick and compared with plasma concentrations measured by chromogenic substrate assays.
- Logistic regression analysis identified optimal plasma concentration thresholds for DXI and DTI detection.
Main Results:
- DOAC Dipstick demonstrated 100% NPV and sensitivity for excluding DOACs at plasma concentrations ≥30 ng/mL.
- At optimized thresholds (≥14 ng/mL for DXI, ≥19 ng/mL for DTI), diagnostic performance improved significantly, with values ranging from 86% to 100%.
- Visual DOAC Dipstick analysis showed 100% agreement with the optoelectronic DOASENSE Reader.
Conclusions:
- DOAC Dipstick urine testing is a reliable method to exclude the presence of DOACs at plasma concentrations >14 ng/mL for DXIs and >19 ng/mL for DTIs.
- The study supports the use of DOAC Dipstick as a valuable tool for rapid assessment of DOAC presence in clinical practice.
- Further studies are needed to confirm its performance in detecting lower DOAC plasma concentrations.