Low-radiation dose scan protocol for preoperative imaging for dental implant surgery using deep learning-based
Yuki Sakai1, Erina Kitamoto2, Kazutoshi Okamura2
1Division of Radiology, Department of Medical Technology, Kyushu University Hospital, 3-1-1, Maidashi, Higashi-ku, Fukuoka, Japan. yuki.sakai.san@gmail.com.
Objectives:
This study aimed to investigate the impact of a deep learning-based reconstruction (DLR) technique on image quality and reduction of radiation exposure, and to propose a low-dose multidetector-row computed tomography (MDCT) scan protocol for preoperative imaging for dental implant surgery.
Methods:
The PB-1 phantom and a Catphan phantom 600 were scanned using volumetric scanning with a 320-row MDCT scanner. All scans were performed with a tube voltage of 120 kV, and the tube current varied from 120 to 60 to 40 to 30 mA. Images of the mandible were reconstructed using DLR. Additionally, images acquired with the 120-mA protocol were reconstructed using filtered back projection as a reference. Two observers independently graded the image quality of the mandible images using a 4-point scale (4, superior to reference; 1, unacceptable). The system performance function (SPF) was calculated to comprehensively evaluate image quality. The Wilcoxon signed-rank test was employed for statistical analysis, with statistical significance set at p value < 0.05.
Results:
There was no significant difference between the image quality acquired with the 40-mA tube current and reconstructed with the DLR technique (40DLR), and that acquired with the reference protocol (3.00, 3.00, p = 1.00). The SPF at 1.0 cycles/mm acquired with 40DLR was improved by 156.7% compared to that acquired with the reference protocol.
Conclusions:
Our proposed protocol, which achieves a two-thirds reduction in radiation dose, can provide a minimally invasive MDCT scan of acceptable image quality for dental implant surgery.