Ultrasound-based method for individualized estimation of central aortic blood pressure from flow velocity and
Shuran Zhou1, Yang Yao2, Wenyan Liu1
1College of Medicine and Biological Information Engineering, Northeastern University, Shenyang, 110169, China.
Abstract:
Central aortic blood pressure (CABP) is a better predictor for cardiovascular events than brachial blood pressure. However, direct CABP measurement is invasive. The objective of this paper is to develop an ultrasound-based method using individualized Windkessel (WK) models for non-invasive estimation of CABP. Three WK models (with two-, three- and four-element WK, named, WK2, WK3 and WK4, respectively) were created and the model parameters were individualized based on aortic flow velocity and diameter waveforms measured by ultrasound (US). Experimental data were acquired in 42 subjects aged 21-67 years. The CABP estimated by WK models was compared with the reference CABP obtained using a commercial system. The results showed that the overall performance of the WK3 and WK4 models was similar, outperforming the WK2 model. The estimated CABP based on WK3/WK4 model showed good agreement with the reference CABP: the absolute errors of systolic blood pressure (SBP), 2.4 ± 2.1/2.4 ± 2.0 mmHg; diastolic blood pressure (DBP), 1.4 ± 1.1/1.7 ± 1.5 mmHg; mean blood pressure (MBP), 1.3 ± 0.8/1.3 ± 0.8 mmHg; pulse pressure (PP), 3.0 ± 2.3/3.2 ± 2.6 mmHg; the root mean square error (RMSE) of the waveforms, 2.5 ± 1.0/2.6 ± 1.1 mmHg. Therefore, the proposed method can provide a non-invasive CABP estimation during routine cardiac US examination.
More Related Videos
Related Concept Videos
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(two-step method)
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.