Comparison of Toric Intraocular Lens Alignment Using Image-Guided System and Manual Marking Technique
1Department of Ophthalmology, Aritmi Osmangazi Hospital, Bursa, Turkey.
The Callisto eye image-guided system is as effective as manual marking for positioning toric intraocular lenses (IOLs), demonstrating comparable visual acuity outcomes in cataract surgery patients.
Area of Science:
- Ophthalmology
- Surgical Technology
- Medical Devices
Background:
- Toric intraocular lenses (IOLs) correct astigmatism during cataract surgery.
- Accurate IOL positioning is crucial for effective astigmatism correction.
- Traditional manual marking techniques have been the standard for IOL alignment.
Purpose of the Study:
- To compare the accuracy of the Callisto eye image-guided system against manual marking for toric IOL placement.
- To evaluate the impact of each positioning method on visual outcomes.
Main Methods:
- Eighty patients received toric IOLs, randomized into an image-guided group (Callisto eye) or a manual marking group.
- Preoperative and 3-month postoperative assessments included visual acuity, refraction, and astigmatism.
- Toric IOL alignment was measured by the deviation from the intended axis.
Main Results:
- Preoperative characteristics were similar between the image-guided and manual marking groups.
- No significant differences were observed in uncorrected or corrected distance visual acuity at 3 months.
- The mean deviation from the intended axis was 2.04±1.84 degrees for the Callisto eye group and 3.24±2.64 degrees for the manual marking group, with no significant impact on visual acuity.
Conclusions:
- The Callisto eye image-guided system provides comparable accuracy to manual marking for toric IOL positioning.
- Image-guided systems offer a viable alternative to manual marking, maintaining effective astigmatism correction and visual outcomes.