Optical Coherence Tomography Based Biomechanical Fluid-Structure Interaction Analysis of Coronary Atherosclerosis Progression
Published on: January 15, 2022
Assessment of Image Quality of Coronary Computed Tomography Angiography in Obese Patients by Comparing Deep Learning
Hongwei Wang1, Rui Wang1, Ying Li1
1From the Department of Radiology, Beijing Anzhen Hospital, Capital Medical University.
Objective:
The aim of the study was to evaluate the image quality of coronary computed tomography (CT) angiography (CCTA) in obese patients by using deep learning image reconstruction (DLIR) in comparison with adaptive statistical iterative reconstruction Veo (ASiR-V).
Methods:
We prospectively evaluated 60 obese patients (body mass index [BMI] ≥ 30 kg/m2) who underwent coronary CT angiography in a single center. All CT scans were performed with GE Revolution 256-row CT at 120 kV (group A; 20 men, 10 women; mean age = 54.3 years; mean BMI = 33.4 kg/m2) or 100 kV (group B; 18 men; 12 women; mean age = 56.8 years; mean BMI = 32.9 kg/m2). Images in group A were reconstructed using ASiR-V, whereas images in group B were reconstructed using ASiR-V, DLIR-medium (DLIR-M), and DLIR-high (DLIR-H). Three blinded independent readers assessed the subjective image quality and measured the objective image quality. Radiation dose estimates were calculated and compared between patients by using 0.014 and 0.026 mSv·mGy-1 cm-1 corresponding to chest and heart conversion coefficients, respectively.
Results:
The subjective score was significantly higher for images reconstructed using 120-kV ASiR-V (3.8), DLIR-M (3.9), and DLIR-H (4.0) compared with those reconstructed using 100-kV ASiR-V (3.5). Image noise was significantly lower in images reconstructed using DLIR-H compared with those reconstructed using other reconstruction algorithm (P < 0.001, respectively). The contrast-to-noise ratio was significantly higher in the DLIR-H group than in the groups using other reconstruction algorithm (P < 0.001). The effective radiation dose was significantly lower in group B than in group A (P < 0.001).
Conclusions:
Compared with ASiR-V, DLIR improved image quality in obese individuals without comprising image quality or increasing the radiation dose.
More Related Videos
09:57Development and Evaluation of 3D-Printed Cardiovascular Phantoms for Interventional Planning and Training
Published on: January 18, 2021
09:32Time-Resolved, Dynamic Computed Tomography Angiography for Characterization of Aortic Endoleaks and Treatment Guidance via 2D-3D Fusion-Imaging
Published on: December 9, 2021
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion,...