Swallowing assessment in patients with dysphagia: Validity and reliability of a pocket-sized ultrasound system
Katharina Winiker1,2,3, Rebecca Hammond1,2,4, Paige Thomas1,2,4
1Department of Psychology, Speech and Hearing, University of Canterbury, Christchurch, New Zealand.
Background:
The use of ultrasound as an adjunct to clinical swallowing evaluation provides quantitative physiological and morphological data. As a low-risk procedure, ultrasound imaging can be performed outside of a medical setting. This is particularly important for patients living in rural areas with restricted access to a hospital. Technical advances have produced pocket-sized ultrasound technology that is more affordable, and therefore within the fiscal reach of most allied health services.
Aims:
To explore the validity and reliability of pocket-sized ultrasound technology in dysphagia assessment.
Methods & Procedures:
Data were acquired from 43 patients with dysphagia using the Clarius ultrasound device. Ultrasound and videofluoroscopic measures of hyoid and laryngeal displacement during liquid and puree swallowing were collected concurrently to quantify correlation and agreement between identical measures derived from the two instruments. Reliability of ultrasound was assessed for measures of hyoid and laryngeal displacement, tongue thickness, and size of the submental muscles in eight patients. Reliability was evaluated for the entire process of data acquisition including scanning and online measurement using an iPad in a clinical setting and for offline measurement on a computer screen to explore environmental influences on reliability.
Outcomes & Results:
Results revealed poor correlation between the measures of interest across instruments. Reliability of the entire process of data acquisition in a clinical setting was insufficient while reliability was more promising for offline measurements.
Conclusions & Implications:
The clinical use of pocket-sized ultrasound devices, such as the Clarius system, for swallowing evaluation is not indicated at this time. Enhanced validity and reliability of the entire process of data acquisition are needed prior to clinical translation of such technology.
What This Paper Adds:
What is already known on the subject The use of ultrasound allows for radiation-free, non-invasive swallowing assessment. Some data suggest that ultrasound is valid and reliable in the evaluation of swallowing using standard-sized equipment. Insufficient validity and reliability have been reported for pocket-sized ultrasound technology in the assessment of healthy swallowing. What this paper adds to existing knowledge This research is the first to provide validity and reliability data of the pocket-sized Clarius technology in the evaluation of swallowing in patients with dysphagia. Insufficient validity and reliability of online data acquisition in a clinical environment were found. Reliability for offline measurement was more promising. What are the potential or actual clinical implications of this work? The clinical use of pocket-sized ultrasound devices, such as the Clarius system, for swallowing assessment is not indicated at this time.
Related Concept Videos
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Upper GI Series: Barium Swallow
Purpose and Procedure
Patients undergoing this procedure ingest a liquid containing barium sulfate with a chalky...
Ultrasonography
During an ultrasonography procedure, a handheld device called...
Ultrasound I: Abdominal Ultrasonography
Abdominal ultrasonography, commonly known as abdominal ultrasound, is a vital, non-invasive medical imaging technique widely used in healthcare.
Procedure:
This diagnostic tool allows the clinician to visually inspect internal structures within the abdomen, including vital organs such as the liver, gallbladder, pancreas, kidneys, and spleen.
The abdominal ultrasound process begins with applying a special gel to the patient's skin over the abdomen. This gel enhances the...