Dynamic Monitoring of Seroconversion using a Multianalyte Immunobead Assay for Covid-19
Published on: February 16, 2022
Comparison of six antibody assays and two combination assays for COVID-19
Masaki Yamamoto1,2, Kazuyuki Okazaki3, Yoko Kitai3
1Department of Clinical Laboratory Medicine, Kyoto University Graduate School of Medicine, 54 Shogoin-Kawaharacho, Sakyo-ku, Kyoto, 606-8507, Japan. masakiy@kuhp.kyoto-u.ac.jp.
Introduction:
In this work, six SARS-CoV-2-specific antibody assays were evaluated, namely, two pan-immunoglobulin (pan-Ig) assays [Roche Elecsys Anti-SARS-CoV-2 (named "Elecsys" in this study) and the PerkinElmer SuperFlex™ Anti-SARS-CoV-2 Ab Assay (SuperFlex_Ab)], two IgM assays [SuperFlex™ Anti-SARS-CoV-2 IgM Assay (SuperFlex_IgM) and YHLO iFlash-SARS-CoV-2 IgM (iFlash_IgM)], and two IgG assays [SuperFlex™ Anti-SARS-CoV-2 IgG Assay (SuperFlex_IgG) and iFlash-SARS-CoV-2 IgG (iFlash_IgG)]. Combination assays of SuperFlex™ (SuperFlex_any) and iFlash (iFlash_any) were also evaluated.
Methods:
A total of 438 residual serum samples from 54 COVID-19 patients in the COVID-19 group and 100 samples from individuals without evidence of SARS-CoV-2 infection in the negative control group were evaluated.
Results:
In the early stage of COVID-19 infection, within 14 days of symptom onset, the seropositive rate was lower than that of the late stage 15 days after onset (65.4% vs 99.6%). In the total period, the pan-Ig and IgG assays had higher sensitivity (90.8-95.3%) than the IgM assays (36.5-40.7%). SuperFlex_Ab and SuperFlex_any had higher sensitivity than Elecsys and SuperFlex_IgG (p < 0.05). The specificity of all the assays was 100%, except for SuperFlex_IgM (99.0%). The concordance rate between each assay was higher (96.4-100%) in the late stage than in the early stage (77.4-98.1%).
Conclusion:
For the purpose of COVID-19 diagnosis, antibody testing should be performed 15 days after onset. For the purpose of epidemiological surveillance, highly sensitive assays should be used as much as possible, such as SuperFlex_Ab, iFlash_IgG and their combination. IgM assays were not suitable for these purposes.