Preliminary Experience of 5.0 T Higher Field Abdominal Diffusion-Weighted MRI: Agreement of Apparent Diffusion
Yunfei Zhang1,2, Chun Yang1,3, Liang Liang3
1Shanghai Institute of Medical Imaging, Fudan University, Shanghai, China.
Background:
Recently, a prototype 5.0 T whole-body MRI scanner was developed. A 5.0 T diffusion-weighted imaging (DWI) may help overcome the issues that limit 3.0 T DWI.
Purpose:
To evaluate the feasibility of 5.0 T high-field DWI in the upper abdomen and assess the agreement of the apparent diffusion coefficient (ADC) with that from 3.0 T abdominal DWI.
Study Type:
Prospective proof of concept.
Population:
Nine volunteers (mean ± SD age: 37.3 ± 7.0 years, 8 M), eight healthy and one with liver and kidney cysts.
Field Strength/Sequence:
3.0 T and 5.0 T; respiratory-triggered spin-echo echo-planar-imaging (SE-EPI)-based DWI sequence.
Assessment:
Subjective image quality scores. The ADC values in abdominal organs (liver, pancreas, spleen, and kidney) were measured by two observers for evaluating the interobserver and interfield agreement.
Statistical Tests:
Wilcoxon-rank sum test, Bland-Altman analysis, intraclass correlation coefficients (ICCs), and coefficients of variation (CVs).
Results:
The 5.0 T DWI displayed an increase in subjective image quality score compared to 3.0 T DWI without the significant difference (3.0 T DWI: 3.50 ± 0.47, 5.0 T DWI: 3.72 ± 0.42, P = 0.157). Both the interfield and interobserver agreements of ADC values were substantial to excellent (ICCs = 0.640-0.902). For all four upper abdominal organs, there were no significant differences between the ADC values measured by two observers and between the ADC values of 3.0 T and 5.0 T DWI (P = 0.134-1.000). The CVs of ADC measurements from 3.0 T and 5.0 T DWI were all less than 15.0% (6.7%-14.2%).
Data Conclusion:
The substantial to excellent agreements between the ADC values measured with 3.0 T and 5.0 T DWI for liver, pancreas, spleen, and kidney suggested that 5.0 T DWI can be applied for abdominal imaging. The ADC values from 5.0 T abdominal DWI hold the potential to serve as the quantitative markers for clinical investigations.
Evidence Level: