Related Experiment Video
Updated: Oct 4, 2025

Monitoring Lung Function with Electrical Impedance Tomography in the Intensive Care Unit
Published on: September 6, 2024
Lung Ultrasound and Electrical Impedance Tomography During Ventilator-Induced Lung Injury
Irene Steinberg1, Iacopo Pasticci1, Mattia Busana1
1Department of Anesthesiology, Intensive Care and Emergency Medicine, University of Göttingen, Göttingen, Germany.
Objectives:
Lung damage during mechanical ventilation involves lung volume and alveolar water content, and lung ultrasound (LUS) and electrical impedance tomography changes are related to these variables. We investigated whether these techniques may detect any signal modification during the development of ventilator-induced lung injury (VILI).
Design:
Experimental animal study.
Setting:
Experimental Department of a University Hospital.
Subjects:
Forty-two female pigs (24.2 ± 2.0 kg).
Interventions:
The animals were randomized into three groups (n = 14): high tidal volume (TV) (mean TV, 803.0 ± 121.7 mL), high respiratory rate (RR) (mean RR, 40.3 ± 1.1 beats/min), and high positive-end-expiratory pressure (PEEP) (mean PEEP, 24.0 ± 1.1 cm H2O). The study lasted 48 hours. At baseline and at 30 minutes, and subsequently every 6 hours, we recorded extravascular lung water, end-expiratory lung volume, lung strain, respiratory mechanics, hemodynamics, and gas exchange. At the same time-point, end-expiratory impedance was recorded relatively to the baseline. LUS was assessed every 12 hours in 12 fields, each scoring from 0 (presence of A-lines) to 3 (consolidation).
Measurements And Main Results:
In a multiple regression model, the ratio between extravascular lung water and end-expiratory lung volume was significantly associated with the LUS total score (p < 0.002; adjusted R2, 0.21). The variables independently associated with the end-expiratory difference in lung impedance were lung strain (p < 0.001; adjusted R2, 0.18) and extravascular lung water (p < 0.001; adjusted R2, 0.11).
Conclusions:
Data suggest as follows. First, what determines the LUS score is the ratio between water and gas and not water alone. Therefore, caution is needed when an improvement of LUS score follows a variation of the lung gas content, as after a PEEP increase. Second, what determines the end-expiratory difference in lung impedance is the strain level that may disrupt the intercellular junction, therefore altering lung impedance. In addition, the increase in extravascular lung water during VILI development contributed to the observed decrease in impedance.
More Related Videos
05:51Author Spotlight: Unraveling the Impact of Mechanical Ventilation on Diaphragm Function and Patient Outcomes
Published on: November 3, 2023
06:22Surfactant Depletion Combined with Injurious Ventilation Results in a Reproducible Model of the Acute Respiratory Distress Syndrome ARDS
Published on: April 7, 2021
Related Concept Videos
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...