Reliability of an Ion-Selective Electrode as a Simple Diagnostic Tool for Mastitis
Kazuna Furukawa1, Katsumi Mizuno1, Minami Azuma1
1Department of Pediatrics, Showa University School of Medicine, Shinagawa-ku, Tokyo, Japan.
Background:
Developing a simple quantitative tool for mastitis diagnosis is essential. The Ion-Selective Electrode for sodium has been reported to reliably measure sodium concentrations in human milk.
Research Aim:
To determine whether an Ion-Selective Electrode measurement of sodium:potassium ratios could serve as a diagnostic tool for mastitis and, if so, to determine the diagnostic cut-off value.
Methods:
A total of 107 milk samples, including 55 from milk bank donors and 52 from participants with mastitis, were studied. The sodium:potassium ratios were determined in 33 samples (without mastitis n = 15; with mastitis n = 18) by the Ion-Selective Electrode and ion chromatography. The remaining 74 samples (donor milk n = 40; participants with mastitis n = 34) were analyzed by Ion-Selective Electrode only. Values were averaged over three measurements for each method.
Results:
The median postpartum months of donors and participants with mastitis were 2 and 3 months, respectively. The mean (SD) sodium:potassium ratios without and with mastitis were 0.5 (0.1) and 1.7 (1.2), respectively. A positive correlation existed between sodium:potassium ratios obtained from the two methods (r = 0.98). Area under the curve values were 0.951 (95% CI [0.904, 0.986]) for the Ion-Selective Electrode (N = 107) and 0.978 (95% CI [0.926, 1.000]) for the ion chromatography (n = 33) methods. The optimal cut-off value for the Ion-Selective Electrode method was 0.60, with 86.5% sensitivity and 92.7% specificity.
Conclusions:
The Ion-Selective Electrode was sufficiently accurate for the diagnosis of mastitis. Cohort studies are needed to explore the relationship between sodium:potassium ratios and clinical outcomes.
Related Concept Videos
Potentiometry: Membrane Electrodes
Ion-Exchange Chromatography
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a...