Evaluation of Breaking Force of Different Suture Materials Used in Dentistry: An In Vitro Mechanical Comparison
Mattia Manfredini1, Susanna Ferrario1, Paola Beretta1
1Maxillofacial Surgery and Odontostomatology Unit, Implant Center for Edentulism and Jawbone Atrophies, Fondazione IRCCS Cà Granda Ospedale Maggiore Policlinico, University of Milan, Via della Commenda 10, 20122 Milan, Italy.
Abstract:
The success of surgical procedures is strictly related to the biomechanical properties of the suture. Mechanical comparisons are scarcely reported in the literature, so the purpose of the present study was to evaluate and compare the mechanical behavior of different sutures commonly used in oral surgery in terms of traction resistance. Sutures made of eight different materials were analyzed: silk (S), polyglycolide-co-caprolactone (PGCL), polypropylene (PP), rapid polyglycolide (rPGA), standard polyglycolide (PGA), polyamide (PA), polyester (PE), and polyvinylidene fluoride (PVDF). For each material, three different sizes were tested: 3-0, 4-0, and 5-0. The breaking force of each suture was assessed with a uniaxial testing machine after being immersed in artificial saliva at 37 °C. The outcomes analyzed were the breaking force, the needle-thread detachment breaking-point and the node response after forward-reverse-forward (FRF) tying when subjected to a tensile force. The 3-0 rPGA provided the maximum resistance, while the lowest value was recorded for the 5-0 PGCL. In general, 3-0 and 4-0 gauges showed non-statistically significant differences in terms of needle-thread detachment. The highest needle-thread detachment was found for the 3-0 PGA, whereas the lowest value was observed for the 5-0 PGCL. After tying the knot with an FRF configuration, the thread that showed the highest resistance to tension was the 3/0 silk, while the thread with the lowest resistance was the 5/0 silk. These data should be considered so that the operator is aware of as many aspects as possible on the behavior of various materials to ensure successful healing.