Related Experiment Video
Updated: Oct 3, 2025

Transthoracic Speckle Tracking Echocardiography for the Quantitative Assessment of Left Ventricular Myocardial Deformation
Published on: October 20, 2016
Semiautomated Software to Improve Stability and Reduce Operator-Induced Variation in Vascular Ultrasound Speckle
Nirmala Rajaram1,2, Brian J Thelen1,3,4, James D Hamilton1
1Veterans Affairs Ann Arbor Healthcare System, Ann Arbor, Michigan, USA.
This study developed semiautomated ultrasound software to measure arterial wall distensibility, reducing user variability in predicting arteriovenous fistula maturation for hemodialysis patients.
Area of Science:
- Vascular Ultrasound
- Medical Imaging Analysis
- Nephrology
Background:
- Arteriovenous fistula (AVF) maturation is critical for hemodialysis in end-stage renal disease (ESRD) patients.
- Ultrasound is a key imaging modality for assessing AVF maturation.
- Accurate measurement of vascular distensibility is essential for predicting AVF success.
Purpose of the Study:
- To develop and evaluate ultrasound software for measuring circumferential vessel wall strain (distensibility).
- To assess and reduce user-induced variability in distensibility measurements from Digital Imaging and Communications in Medicine (DICOM) data.
- To improve the reliability of distensibility measurements for predicting AVF maturation.
Main Methods:
- Ultrasound scans of arteries from 10 subjects undergoing AVF surgery were analyzed.
- The Kanade-Lucas-Tomasi (KLT) feature-tracking algorithm was used to track vessel walls in DICOM cine loops.
- Two semiautomated measurement methods were developed and compared to manual user measurements.
Main Results:
- Manual point selection for vessel tracking showed significant user variability (up to 9.1 mm).
- This variability led to a wide range in distensibility measurements (5.63% to 41.04%).
- Semiautomated methods significantly reduced measurement variation and correlated well with median user values.
Conclusions:
- Standardizing point selection through semiautomated software enhances reproducibility and reliability of distensibility measurements.
- The developed software can aid clinical studies in understanding vascular compliance's role in fistula maturation.
- Improved measurement accuracy may lead to better prediction of AVF maturation for hemodialysis patients.