Continuous Monitoring of Blood Pressure Using a Wrist-Worn Cuffless Device
Gabriel Sayer1, Greta Piper2, Esther Vorovich3
1Department of Medicine, Columbia University Irving Medical Center, New York, New York, USA.
Background:
Measuring an accurate blood pressure (BP) in ambulatory patients is critical for therapeutic decisions. An accurate noninvasive device that measures BP continuously could provide a more comprehensive assessment of BP. The aim of this study was to determine whether the LiveOne device, a cuffless wristband that provides continuous BP measurements, is accurate in comparison to an invasive arterial line (A-line).
Methods:
We enrolled hospitalized patients and healthy volunteers in this multicenter study. All patients had an invasive A-line. Waveforms were simultaneously downloaded from the A-line and LiveOne device. The primary outcome was the correlation between the LiveOne device and the A-line for systolic BP (SBP) and diastolic BP (DBP).
Results:
Thirty-four patients with 233 measurement periods were included. Mean age was 60.7 ± 15.2 years, and 15 (44%) were female. The correlation for SBP was 0.91 and for DBP was 0.85. The mean band error was 0.0 ± 6.9 mm Hg for SBP and 1.2 ± 5.7 mm Hg for DBP. The mean absolute error was 8.2 ± 5.8 mm Hg for SBP and 6.4 ± 3.9 mm Hg for DBP. For SBP, 98% of LiveOne measurements were within 15 mm Hg and for DBP, 92% of LiveOne measurements were within 10 mm Hg.
Conclusions:
The LiveOne device provides continuous, noninvasive BP measurements that are accurate in comparison to A-line measurements. The portability and unobtrusive nature of this device and the ability to provide continuous BP measurements may offer advantages over currently available BP monitors.
Clinical Trials Registration:
Trial Number NCT03919136.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessment of blood pressure in brachial artery(two-step method)
Pre-Procedural Guidelines for Assessing Blood Pressure