Comparison of Agreement and Accuracy using Binocular Wavefront Optometer with Autorefractor and Phoropter
Published on: September 16, 2025
Intraobserver Repeatability of Corneal and Anterior Segment Parameters Obtained Using a Scheimpflug Camera-Placido
1Department of Ophthalmology, Siirt Kurtalan State Hospital, Siirt, Turkey.
Objectives:
The aim of this study was to assess the intraobserver repeatability of central corneal thickness, peripheral corneal thickness (PCT), keratometry values (steep K, flat K), white-to-white diameter, and anterior chamber depth using the Sirius topography device (Costruzione Strumenti Oftalmici, Florence, Italy) in healthy eyes.
Methods:
A Sirius device was used by a single examiner to assess 100 eyes in 50 healthy patients. Two consecutive scans were acquired for each eye. Repeatability was assessed using test-retest variability, the coefficient of variation (COV), and the intraclass correlation coefficient (ICC).
Results:
Fifty patients (100 eyes) met the inclusion criteria. There were 18 women (36%) and 32 men (64%), with an age range of 23 to 56 years. The mean age was 30.38±8.03 years. A COV of 0.4% or less and an ICC of more than 0.99 (showing excellent repeatability) were achieved for most parameters, with the exception of PCT (at 2.5-mm temporal, superior, inferior, and nasal thickness).
Conclusion:
The anterior segment parameters obtained using the Sirius Scheimpflug camera- Placido corneal topography system were highly repeatable.