Image quality assessment of artificial intelligence iterative reconstruction for low dose aortic CTA: A feasibility
Wanjiang Li1, Yongchun You1, Sihua Zhong2
1Department of Radiology, West China Hospital of Sichuan University, Chengdu, Sichuan, China.
Purpose:
To investigate the image quality and feasibility of a novel artificial intelligence iterative reconstruction (AIIR) algorithm for aortic computer tomography angiography (CTA) with a low radiation dose and contrast material (CM) dosage protocol in comparison with hybrid iterative reconstruction (HIR) algorithm for standard-of-care aortic CTA.
Methods:
Fifty consecutive patients (mean age 58 ± 14 years, mean BMI 24.5 ± 4.7 kg/m2) with aortic diseases were prospectively enrolled. All patients underwent at least twice follow-up aortic CTA examinations. Standard dose CT (SDCT) was applied in the initial follow-up examination (100 kVp, auto mAs, contrast dose 0.8 mgL/kg), images were reconstructed with HIR (SDCT-HIR). In the second follow-up examination, patients underwent scanning with low dose CT (LDCT) (70 kVp, auto mAs, contrast dose 0.5 mgL/kg), images were reconstructed with HIR (LDCT-HIR) as well as AIIR (LDCT-AIIR). Attenuation values, noise, signal-to-noise ratio (SNR), and contrast-to-noise ratio (CNR) were measured for objective analysis. Subjective image quality was rated by two blinded radiologists using a 5-point scale. The effective radiation dose and CM dosage were also recorded.
Results:
The effective radiation dose (1.58 ± 0.17 mSv vs. 9.96 ± 1.05 mSv, P < 0.001) and CM dosage (34.38 ± 5.43 ml vs. 54.64 ± 8.63 ml, P < 0.001) achieved a remarkable reduction of 84.14% and 37.08% in the LDCT compared to the SDCT. The attenuation was similar among the three reconstructed images (P > 0.05). Compared to LDCT-HIR images, LDCT-AIIR showed a lower noise and higher SNR and CNR. For qualitative analysis, there were no significant differences between the LDCT-AIIR and the SDCT-HIR images among four metrics (P > 0.05).
Conclusions:
Compared to standard-of-care aortic CTA with HIR, the application of the AIIR algorithm allows for radiation dose and CM dosage reduction while preserving image quality on low dose aortic CTA.
More Related Videos
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...