Related Experiment Video
Updated: Oct 2, 2025

Implementation of a Reference Interferometer for Nanodetection
Published on: April 26, 2014
Homodyne laser vibrometer modified by an LCVR for measurement at the nanometer level
Abstract:
The existence of periodic nonlinear error restricts the performance of the homodyne laser vibrometer in sub-fringe amplitude vibration measurement. A homodyne laser vibrometer with nanoscale-amplitude detectability by using a liquid crystal variable retarder (LCVR) is proposed. The LCVR introduces an extra variation of optical path difference larger than the laser wavelength to acquire a full ellipse so that the nonlinearity correction parameters could be pre-extracted. The experiments showed that the nonlinear error could be well suppressed with the correction process based on the pre-extracted parameters, and the detectable minimum amplitude is less than 1 nm. In addition, measurement of vibration with the reflectivity of measured targets down to 0.048% was achieved with an automatic-gain-control module.
More Related Videos
15:25Design and Characterization Methodology for Efficient Wide Range Tunable MEMS Filters
Published on: February 4, 2018
12:00Synthesis of Gold Nanoparticle Integrated Photo-responsive Liposomes and Measurement of Their Microbubble Cavitation upon Pulse Laser Excitation
Published on: February 24, 2016