Software for Analysis of Heart Rate and Blood Pressure Time-series Data from the Valsalva Maneuver
Published on: June 27, 2025
Intrasession Reliability Analysis for Oscillometric Blood Pressure Method Using a Digital Blood Pressure Monitor in
Sabina Barrios-Fernandez1, Eduardo Manuel Sosa-Sánchez2, Jorge Carlos-Vivas2
1Social Impact and Innovation in Health (InHEALTH) Research Group, Faculty of Sport Sciences, University of Extremadura, 10003 Caceres, Spain.
Abstract:
Blood Pressure (BP) is one of the most used measured clinical parameters in health promotion and intervention. BP measures can vary due to different parameters, so we aim to study the intrasession test-retest reliability for an oscillometric method using a digital tensiometer in the Peruvian population aged over 15 with and without a diagnosis of hypertension (HT). Data were taken from the Demographic and Family Health Survey conducted in Peru in 2019. Technicians had to follow a standardized protocol on the conditions to carry out a valid and reliable measurement. Relative reliability was excellent in most cases (intraclass correlation coefficient > 0.9); absolute reliability was excellent (standard error of measurement < 5%) and smallest real difference < 10% in most cases. The Bland-Altman plot showed a systematic error of 2.36 for systolic BP in men and 2.16 in women, and 0.823 for diastolic BP in men and 0.71 for diastolic BP in women. Results suggest that the oscillometric method with a digital blood pressure monitor was reliable in absolute and relative terms in this population, so it could be used as a reliable control test to measure changes after an intervention.
Related Concept Videos
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(two-step method)
Errors occurring during blood pressure monitoring
Several factors...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.