Availability of Home sleep apnea test equipment LS-140 on a comparison with Polysomnography
Yuki Mieno1, Masamichi Hayashi2, Mariko Hirochi1
1Department of Respiratory Medicine Ⅰ, Fujita Health University, School of Medicine, Toyoake, Aichi, Japan.
Objective:
The prevalence of obstructive sleep apnea (OSA) in Japan is 9% among males and 3% among females. Up to 2.5 million patients are estimated to suffer from the disease, but limited number of facilities are capable of carrying out polysomnography (PSG), leaving more than 80% of these individuals are undiagnosed. In recent years, the development of new portable sleep monitoring (PMs) devices has been remarkable. We evaluate the correlation between the results of the LS-140 PMs device (Fukuda Denshi Tech Co. Ltd.), released in 2017, and those of PSG.
Methods:
We obtained contemporaneous data from the same patients by equipping 58 patients with PMs (LS-140) devices while they underwent PSG. Our primary outcome was Case 2 of the intraclass correlation coefficient (ICC), i.e., the ICC (2.1). And we used a Bland-Altman analysis to compare the apnea-hypopnea index (AHI) given by PSG and the respiratory event index (REI) given by LS-140 and examined the sensitivity and specificity of the REI relative to the AHI in the diagnosis of OSA. We also carried out the same comparison but in terms of the presence or absence of periodic limb movements (PLMs).
Results:
The ICC (2.1) between The REI and the AHI was 0.944, a rather high value (p<0.0001). The mean difference between AHI and REI values was -3.6 (p<0.0001), indicating a negative fixed bias. Sensitivity may decrease in groups with PLMs.
Conclusion:
The REI and the AHI are highly correlated, giving LS-140 sufficient diagnostic sensitivity and specificity to screen for OSA.
Related Concept Videos
Sleep Apnea
The condition is more prevalent among...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...