Intra-arterial blood pressure measurement: sources of error and solutions
Suresh Devasahayam1,2, Naveen Gangadharan3,4, C Surekha5
1Department of Bioengineering, Christian Medical College Vellore, 632002, Tamil Nadu, India.
Abstract:
Intra-arterial blood pressure measurement is the cornerstone of hemodynamic monitoring in intensive care units (ICU). Accuracy of the measurement is dependent on the dynamic response of the measuring system, defined by its natural frequency (fnatural) and damping coefficient (Zdamping). Gardner's plot (1981) has long been the only way to determine the accuracy of the pressure measurement. Specific objectives: (i) estimation of the amplitude of error in pressure measurement through simulations based on real-world data, (ii) a novel method to correct the error. Simulated blood pressure waveforms of various heart rates were passed through simulated measurement systems with varying fnatural and Zdamping. The numerical errors in systolic and diastolic pressures and mean error in the measured pressure were used to generate heat maps for the various recording conditions, in the same plot as that by Gardner (1981). fnatural and Zdamping from 121 patient recordings are plotted on these heat maps to demonstrate the fraction of unacceptable recordings. Performance of a tunable filter to correct the error is demonstrated. In many clinical settings, the measurement of intra-arterial pressure is prone to significant error. The proposed tunable filter is shown to improve the accuracy of intra-arterial pressure recording.
Related Concept Videos
Errors occurring during blood pressure monitoring
Several factors...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(two-step method)
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Pre-Procedural Guidelines for Assessing Blood Pressure