Stability with a Constrained Posterior Stabilized Primary Total Knee Arthroplasty Does Not Compromise Durability
Nicholas Yohe1, Douglas J Vanderbrook1, Alain E Sherman1
1Department of Orthopaedic Surgery, Lenox Hill Hospital, Northwell Health, New York.
The Journal of Knee Surgery
|March 3, 2022
Summary
Total knee arthroplasty using a midlevel constraint (MLC) prosthesis effectively improved range of motion and alignment in complex knee deformities. This study found no evidence of loosening or instability, indicating satisfactory outcomes for TKA patients.
Area of Science:
- Orthopedic Surgery
- Biomedical Engineering
Background:
- Total knee arthroplasty (TKA) can be challenging in patients with complex knee deformities, such as ligamentous laxity or posttraumatic arthritis.
- Traditional prostheses may lack the necessary varus-valgus stability for these complex cases.
- Midlevel constraint (MLC) prostheses offer enhanced stability for such challenging TKA procedures.
Purpose of the Study:
- To evaluate the clinical and radiographic outcomes of primary total knee arthroplasty (TKA) utilizing a midlevel constraint (MLC) prosthesis.
- To assess the effectiveness of MLC implants in managing complex knee deformities.
- To determine the incidence of complications such as instability, loosening, or the need for revision surgery.
Main Methods:
- A retrospective review was conducted on 53 patients (62 knees) who underwent primary TKA with an MLC implant.
- Patients were followed for a mean of 31.6 months, with assessment of preoperative and postoperative range of motion (ROM).
- Radiographic review included alignment, radiolucent lines, component loosening, and osteolysis. Outcomes were compared between knees with and without short tibial stem extensions.
Main Results:
- Mean ROM significantly improved from 114.1 degrees preoperatively to 123.5 degrees postoperatively (p=0.001).
- Surgical correction of preoperative varus or valgus deformities resulted in a mean postoperative alignment of 4.3 degrees valgus (p<0.001).
- No radiographic evidence of radiolucent lines, component loosening, or osteolysis was observed. No patients required revision surgery. Arthrofibrosis requiring manipulation under anesthesia occurred in 9.7% of patients.
Conclusions:
- Total knee arthroplasty with a midlevel constraint (MLC) prosthesis yields satisfactory clinical and radiographic results in complex knee deformities.
- The MLC prosthesis demonstrates no evidence of component loosening, osteolysis, or instability at a minimum 2-year follow-up.
- The use of short tibial stem extensions did not significantly impact ROM, alignment, or the incidence of arthrofibrosis compared to non-stemmed implants.
Related Concept Videos
Stability of structures
263
In mechanical engineering, the stability of systems under various forces is critical for designing durable and efficient structures. One fundamental way to explore these concepts is by analyzing systems like two rods connected at a pivot point, O, with a torsional spring of spring constant k at the pivot point. This system is similar in appearance to a scissor jack used to change tires on a car. In this case, the arms of the linkage (equivalent to the rods in this system) are entirely vertical,...
263
Knee Joint
2.4K
The knee joint is the most complicated joint in the body. It consists of three articulations– two tibiofemoral and one patellofemoral. As is characteristic of synovial joints, the knee joint has a thin articular capsule that partially surrounds this joint cavity. Additionally, several ligaments, muscles, and cartilaginous structures support the movement of the knee.
A total of seven ligaments support the knee joint. The patellar ligament, which is also attached to the quadriceps femoris...
A total of seven ligaments support the knee joint. The patellar ligament, which is also attached to the quadriceps femoris...
2.4K
Stability
200
The time response of a linear time-invariant (LTI) system can be divided into transient and steady-state responses. The transient response represents the system's initial reaction to a change in input and diminishes to zero over time. In contrast, the steady-state response is the behavior that persists after the transient effects have faded.
The stability of an LTI system is determined by the roots of its characteristic equation, known as poles. A system is stable if it produces a bounded...
The stability of an LTI system is determined by the roots of its characteristic equation, known as poles. A system is stable if it produces a bounded...
200
Stability of Equilibrium Configuration
552
Understanding the stability of equilibrium configurations is a fundamental part of mechanical engineering. In any system, there are three distinct types of equilibrium: stable, neutral, and unstable.
A stable equilibrium occurs when a system tends to return to its original position when given a small displacement, and the potential energy is at its minimum. An example of a stable equilibrium is when a cantilever beam is fixed at one end and a weight is attached to the other end. If the weight...
A stable equilibrium occurs when a system tends to return to its original position when given a small displacement, and the potential energy is at its minimum. An example of a stable equilibrium is when a cantilever beam is fixed at one end and a weight is attached to the other end. If the weight...
552