Related Experiment Video
Updated: Oct 1, 2025

Implantation of Electroencephalogram and Electrocardiogram Telemetry Devices in Neonatal Rabbit Kits
Published on: February 28, 2025
TonoVet Plus®: Higher reliability and repeatability compared with Tono-Pen XL™ and TonoVet® in rabbits
Sirlene F Barbosa1, Ana Claudia S Raposo1, Francisco de Assis Dórea Neto1
1Veterinary Medicine and Zootechny, Federal University of Bahia, Salvador, Bahia, Brazil.
Objective:
To evaluate the accuracy and repeatability of the Tono-Pen XL™, TonoVet® and TonoVet Plus® tonometers by manometric evaluation, and to establish adjustment equations for intraocular pressure (IOP) estimates in rabbits.
Animal Studied:
Rabbits.
Procedures:
A postmortem study was conducted on seven rabbit eyes to verify the correlation between manometry and tonometry with an artificial incremental increase in IOP from 5 and 60 mmHg. A clinical study was conducted to evaluate accuracy and to establish reference values for the species, with measurement of IOP in 17 animals, for 2 consecutive days, with the same tonometers and calibrations used in the postmortem evaluations.
Results:
There were strong linear trends for all evaluated tonometers. In the in-vivo evaluation, the mean IOP values were: 14.23 ± 1.75 (Tono-Pen XL™); 13.89 ± 2.07 (TonoVet® calibration mode 'd'); 8.88 ± 1.24 (TonoVet calibration mode 'p'); 18.59 ± 1.94 (Tonovet Plus®). There was a significant difference in the two evaluation times for the two TonoVet® calibration modes. The adjustment equations generated from the manometry for the evaluated tonometers were: Y = 0.2570X + 2.219 (Tono-Pen XL™), Y = 0.2289X + 2.389 (TonoVet® 'd'), Y = 0.4043X + 4.062 (TonoVet® 'p'), Y = 0.1233X + 0.3644 (TonoVet Plus®) (X is device-estimated IOP).
Conclusions:
All evaluated tonometers were well correlated with the manometry, with an underestimation of IOP by all devices. Applying adjustment formulas may compensate for systematic errors. TonoVet Plus® was well tolerated, and showed better repeatability and reliability in successive evaluations.