Related Experiment Video
Updated: Oct 1, 2025

A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
Electrocardiographic changes during haemodialysis and the potential impact on subcutaneous implantable cardioverter
Benedict M Wiles1, John M Morgan2, Adam Kirk3
1Cardiac Rhythm Management Research Department, University Hospital Southampton NHS Foundation Trust, Southampton, United Kingdom; Faculty of Medicine, University of Southampton, Southampton, United Kingdom.
Insights
Subcutaneous implantable cardioverter-defibrillator (S-ICD) eligibility in hemodialysis patients can change during dialysis due to R:T wave variations. Pre-dialysis screening is recommended to ensure continued S-ICD eligibility.
Area of Science:
- Cardiology
- Biomedical Engineering
Background:
- Hemodialysis patients needing defibrillator therapy may benefit from subcutaneous implantable cardioverter-defibrillators (S-ICDs).
- Hemodialysis causes dynamic ECG changes (R and T waves) that can affect S-ICD eligibility.
- Continuous assessment of S-ICD eligibility during hemodialysis has not been previously studied.
Purpose of the Study:
- To continuously assess S-ICD eligibility during hemodialysis.
- To identify variations in S-ICD eligibility and their underlying mechanisms.
Main Methods:
- Continuous surface ECG recordings from hemodialysis patients without ICD indication.
- Retrospective automated vector screening at one-minute intervals.
- Calculation of S-ICD eligibility variations and identification of underlying causes.
Main Results:
- 72 vector recordings from 24 patients were analyzed.
- S-ICD eligibility varied during hemodialysis, with 8.4% of vectors changing eligibility.
- High variability (up to 12.5%) was most commonly caused by R:T ratio changes.
Conclusions:
- Hemodialysis can alter S-ICD eligibility, primarily due to R:T wave variations.
- Pre-hemodialysis S-ICD screening is crucial for identifying vectors that maintain eligibility.
- This study highlights the dynamic nature of S-ICD suitability in hemodialysis patients.
Introduction:
Haemodialysis patients who require defibrillator therapy are expected to benefit from the entirely avascular subcutaneous defibrillator (S-ICD), but haemodialysis is associated with dynamic changes in R and T wave amplitude which can impact S-ICD eligibility. A continuous assessment of S-ICD eligibility during haemodialysis has not previously been performed.
Material And Methods:
Continuous surface ECG recordings were obtained from a cohort of patients undergoing maintenance haemodialysis, but without an indication for an ICD. Automated vector screening was retrospectively performed at one-minute intervals throughout the dialysis session. Variations in S-ICD eligibility were calculated and in vectors with high degrees of variation, the underlying mechanism was identified.
Results:
72 vector recordings (mean duration 254.1 ± 6.0 min) were obtained from 24 patients (mean age 64.3 ± 5.5 years, 68% male). At the start of haemodialysis 47 vectors were S-ICD eligible (65.2%). At the end of session, all of these vectors had remained eligible, and an additional 6 vectors had also become eligible (73.6%). High vector score variability was observed in 7 patients and the commonest cause was a progressive change in R:T ratio (71.5%).
Conclusion:
In a haemodialysis population, a single haemodialysis session can be associated with a potential change in S-ICD eligibility in 8.4% of vectors, with up to 12.5% of vectors showing high degrees of variability, most commonly due to variations in R:T ratio. In an S-ICD population with similar characteristics S-ICD screening prior to haemodialysis would be expected to more accurately identify vectors that retain eligibility.
More Related Videos
10:17Real-Time Cardiac Mapping with a Noninvasive Imageless Electrocardiographic Imaging System
Published on: April 11, 2025
18:11A Research Method For Detecting Transient Myocardial Ischemia In Patients With Suspected Acute Coronary Syndrome Using Continuous ST-segment Analysis
Published on: December 28, 2012
Related Concept Videos
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...
Holter Monitor: 24-Hour Monitoring
Cardiomyopathy V: Interprofessional Care
Hemodialysis II: Procedure and Complications
Dysrhythmias V: Evaluating Dysrhythmias
Cardiomyopathy II: Dilated Cardiomyopathy