Left ventricular septal pacing versus left bundle branch pacing in the treatment of atrioventricular block
Yu Zhou1, Jinfeng Wang2, Youquan Wei2
1Department of Emergency, The First Affiliated Hospital of Wannan Medical College, Wuhu, Anhui, China.
Background:
This study aimed to evaluate the feasibility and clinical response of LVSP as an alternative to LBBP.
Methods:
This was a retrospective study of pacemaker implantation, and 46 consecutive patients with pacemaker implantation were enrolled in the study. The patients were divided into the LBBP and LVSP groups. Electrocardiogram characteristics, pacing parameters, cardiac function, and safety events were assessed during implantation and 12-month follow-up.
Results:
The procedure time was significantly increased in the LBBP group compared with the LVSP group (53.52 ± 14.39 min vs. 38.13 ± 11.52 min, respectively, p = .000). The pacing QRS duration (PQRSD) decreased by 14.09 ± 41.80 ms in the LBBP group and increased by 9.70 ± 29.60 ms in the LVSP group (p = .031). Furthermore, the left ventricle activation time (LVAT) was shorter in the LBBP group than in the LVSP group (48.70 ± 13.67 ms vs. 58.70 ± 13.67 ms, p = .032). During the 12-month follow-up, pacing thresholds remained low and stable, and there was no significant decrease in cardiac function. No adverse event was observed during the follow-up period.
Conclusions:
Both LBBP and LVSP are safe and feasible methods. LVSP is a good option when multichannel electrophysiological instruments are not available and when the time available for the procedure is limited.
Related Concept Videos
Dysrhythmias IV: Characteristics of Bradyarrhythmias
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Dysrhythmias VI: Management of Dysrhythmias
Cardiac Catheterization III: Left Heart Catheterization
Heart Failure Drugs: β-Blockers
The Cardiac Cycle
The Process
Electrical signals—sent from the sinoatrial (SA) node in the right atrial wall to the atrioventricular (AV) node between the right atrium and right ventricle—cause both atria to simultaneously contract. When the signal reaches the AV node, it pauses for approximately a tenth of a second, allowing the atria to contract and...