Effect of quantitated stenoses on arterial sound spectrum analysis
American Journal of Surgery
|August 1, 1986
Summary
Spectral analysis of carotid Doppler signals can determine carotid artery stenosis. Percent window, a measurement, accurately differentiates stenosis severity, unaffected by Doppler angle or gain, unlike peak frequency.
Area of Science:
- Vascular Ultrasound
- Medical Imaging
- Hemodynamics
Background:
- Carotid Doppler spectral analysis is a validated method for assessing carotid artery stenosis severity.
- Doppler angle and signal gain are potential confounding factors in spectral analysis accuracy.
Purpose of the Study:
- To evaluate the impact of Doppler angle and signal gain on spectral analysis parameters.
- To determine the reliability of peak frequency, mode frequency, and percent window in quantifying carotid stenosis.
Main Methods:
- An in vitro model using a cadaver artery with varying degrees of artificial stenosis (0-100%) was employed.
- Spectral analysis parameters (peak frequency, mode frequency, percent window) were measured at different Doppler angles (45°, 60°, 75°) and gain settings (low, medium, high).
- Correlation between stenosis percentage and spectral parameters was analyzed.
Main Results:
- Excellent correlation was observed between stenosis percentage and peak frequency (r=0.9520) and percent window (r=-0.9861) at 60° Doppler angle and medium gain.
- Peak frequency and mode frequency decreased at stenoses >82%, mimicking lower stenosis levels.
- Percent window demonstrated minimal overlap across all stenosis degrees and was unaffected by Doppler angle or gain.
Conclusions:
- Percent window is a robust spectral parameter for quantifying carotid artery stenosis, unaffected by Doppler angle or gain.
- Peak frequency is sensitive to Doppler angle and cannot reliably differentiate critical from subcritical stenoses.
- Spectral analysis, particularly percent window, offers a reliable method for stenosis assessment in carotid Doppler studies.