Accuracy of the Data of Biochemical Parameters in Blood Collected Above the Infusion Insertion Site
Shotaro Koike1,2,3,4, Shinsuke Sasaki1,2,3,4, Rika Yano1,2,3,4
1Department of Nursing, Aomori University of Health and Welfare, Hamadate, Aomori, Japan (Dr Koike); Department of Nursing Science, Okayama Prefectural University, Soja-shi, Okayama, Japan (Dr Sasaki); Department of Health Sciences, Hokkaido University, Sapporo, Hokkaido, Japan (Dr Yano).
Abstract:
The authors aimed to investigate the possibility of collecting blood samples from above patient intravenous infusion sites by temporarily stopping the infusions while collecting the blood. A hypotonic infusion solution was administered to 5 male patients through the cephalic veins in the forearms of their left upper limbs. The biochemical data of blood collected from the median cubital veins of both patients' arms were compared. The results showed that infusions could change some biochemical test parameters, such as potassium and glucose, to higher levels, even if the infusion was interrupted temporarily during the blood collection from above the insertion site. Blood counts remained unchanged under the same conditions. If the evaluation is performed solely to assess blood counts, using a blood collection site above the insertion site by temporarily interrupting the infusion is feasible.
Related Concept Videos
Errors occurring during blood pressure monitoring
Several factors...
Drug Concentrations: Measurements
Plasma...
Blood Studies I: ABG and VBG
Arterial Blood Gas (ABG)
Arterial Blood Gas (ABG) studies are crucial for assessing the lungs' ability to supply oxygen and remove carbon dioxide, reflecting the patient's ventilation status. They also help understand the kidneys' capacity to...
One-Compartment Model: IV Infusion
The one-compartment model for IV infusion uses mathematical equations to describe the rate of change in drug quantity in the body. At steady-state or infusion equilibrium, the drug input...