Collarless Taper Slip and Collared Composite Beam Stems Differ in Failure Modes and Reoperation Rates
Richard D Rames1, Anne A Smartt1, Matthew P Abdel1
1Department of Orthopedic Surgery, Mayo Clinic, Rochester, MN.
Comparing two cemented femoral stems in total hip arthroplasty, the Exeter collarless taper slip (CTS) showed higher periprosthetic femoral fracture (PPFX) risk, while the EON composite beam (CB) had increased loosening and intraoperative fracture risks. Surgeons must consider these distinct failure modes.
Area of Science:
- Orthopedic Surgery
- Biomaterials Engineering
Background:
- Cemented femoral components are preferred in elderly patients due to lower risks of periprosthetic fracture and implant loosening.
- This study compares the long-term outcomes of two distinct cemented femoral stem designs.
Purpose of the Study:
- To evaluate and compare survivorship rates free of periprosthetic femoral fracture (PPFX), femoral loosening, all-cause revision, and reoperation between two cemented stem philosophies.
- To identify specific failure modes associated with each stem type in a hybrid total hip arthroplasty cohort.
Main Methods:
- A retrospective single-center study of 1,306 primary hybrid total hip arthroplasties for osteoarthritis (2000-2018).
- Comparison of 798 EON composite beam (CB) stems versus 508 Exeter collarless taper slip (CTS) stems.
- Utilized an inverse treated probability weighted model to control for confounding risk factors.
Main Results:
- No significant difference in overall PPFX risk at 10 years (CB 5% vs. CTS 9%).
- Increased intraoperative fractures in the CB cohort (5%) versus CTS (0%).
- Increased Vancouver B2 PPFX in the CTS cohort (7/508) versus CB (0/798).
- Higher risk of femoral loosening in the CB cohort (6/798) versus CTS (0/508).
- CB cohort demonstrated higher survivorship free of revision (98% vs. 91%) and reoperation (96% vs. 88%) at 10 years.
Conclusions:
- The Exeter collarless taper slip (CTS) stem is associated with an increased risk of periprosthetic femoral fracture requiring revision.
- The EON composite beam (CB) stem shows a higher propensity for femoral component loosening and intraoperative fractures.
- Surgeons should carefully consider the differential failure modes of these cemented stems when selecting an implant design for elderly patients undergoing total hip arthroplasty.
Related Concept Videos
Deformation of a Beam under Transverse Loading
The insights from the bending moment diagram extend to...
Impact Loading on a Cantilever Beam
When an object is dropped onto the free end of a cantilever, its potential energy due to gravity is...
Plastic Deformations
Elastic Curve from the Load Distribution
For all beams, the analysis of the beam's reaction to distributed loads begins by understanding the relationship between a beam's load and the resulting shear forces and bending moments.
Beams
Based on geometry, beams can be straight, tapered, or curved. Straight beams are the most common type and have a constant cross-section throughout their length. Tapered beams, on the other hand, have a varying cross-section along...
Beams with Unsymmetric Loadings
The first moment-area theorem determines the slope at any point on the beam. This theorem indicates that the change in slope between two points on a beam...