Related Experiment Video
Updated: Sep 30, 2025

Author Spotlight: Segmentation and VR for Advanced Neurovascular Interventions
Published on: April 5, 2024
Real-time vessel segmentation and reconstruction for virtual fixtures for an active handheld microneurosurgical
Aravind Venugopal1, Sara Moccia2,3, Simone Foti4
1Department of Computer Science, BITS Pilani, Pilani, Goa, India.
Purpose:
Complications related to vascular damage such as intra-operative bleeding may be avoided during neurosurgical procedures such as petroclival meningioma surgery. To address this and improve the patient's safety, we designed a real-time blood vessel avoidance strategy that enables operation on deformable tissue during petroclival meningioma surgery using Micron, a handheld surgical robotic tool.
Methods:
We integrated real-time intra-operative blood vessel segmentation of brain vasculature using deep learning, with a 3D reconstruction algorithm to obtain the vessel point cloud in real time. We then implemented a virtual-fixture-based strategy that prevented Micron's tooltip from entering a forbidden region around the vessel, thus avoiding damage to it.
Results:
We achieved a median Dice similarity coefficient of 0.97, 0.86, 0.87 and 0.77 on datasets of phantom blood vessels, petrosal vein, internal carotid artery and superficial vessels, respectively. We conducted trials with deformable clay vessel phantoms, keeping the forbidden region 400 [Formula: see text]m outside and 400 [Formula: see text]m inside the vessel. Micron's tip entered the forbidden region with a median penetration of just 8.84 [Formula: see text]m and 9.63 [Formula: see text]m, compared to 148.74 [Formula: see text]m and 117.17 [Formula: see text]m without our strategy, for the former and latter trials, respectively.
Conclusion:
Real-time control of Micron was achieved at 33.3 fps. We achieved improvements in real-time segmentation of brain vasculature from intra-operative images and showed that our approach works even on non-stationary vessel phantoms. The results suggest that by enabling precise, real-time control, we are one step closer to using Micron in real neurosurgical procedures.
More Related Videos
10:25Technical Approach for Infrared Tracking for Soft Tissue Navigation with a Holographic Head-Mounted Display and Preclinical Validation
Published on: September 2, 2025
12:49A Method for 3D Reconstruction and Virtual Reality Analysis of Glial and Neuronal Cells
Published on: September 28, 2019