Related Experiment Video
Updated: Jun 17, 2026

Rat Model of Blood-brain Barrier Disruption to Allow Targeted Neurovascular Therapeutics
Published on: November 30, 2012
Transcarotid transcatheter aortic valve replacement with cerebral embolic protection
Daniel McGrath1, Payam Salehi1, Andrew Weintraub1
1Cardiovascular Center, Tufts Medical Center, Tufts University School of Medicine, Boston, Massachusetts, USA.
Abstract:
There remains concern about the risk of stroke with transcarotid transcatheter aortic valve replacement (TAVR). We describe the use of the Sentinel cerebral protection device with transcarotid TAVR in a patient at high risk for embolic complications. Deployment of the Sentinel filter in the innominate artery accommodated the insertion of the TAVR device through the bovine aortic arch via the contralateral carotid artery. By adapting the "flip-n-flex" technique from our previous transaxillary TAVR experience, we performed left transcarotid TAVR with the Sentinel device in an ergonomic manner and achieved coaxial placement of the transcatheter valve. Inspection of the Sentinel filter after the case revealed captured embolic debris.
More Related Videos
13:10Combined Near-infrared Fluorescent Imaging and Micro-computed Tomography for Directly Visualizing Cerebral Thromboemboli
Published on: September 25, 2016
06:59Improved Registration of 3D CT Angiography with X-ray Fluoroscopy for Image Fusion During Transcatheter Aortic Valve Implantation
Published on: June 3, 2018