Experience with implantation of Select Secure® leads in paediatric patients

Mehmet Küçük1, Sevket Balli2, Huseyin Karadag2

  • 1Department of Paediatric Cardiology, Samsun Training and Research Hospital, Samsun, Turkey.

Acta Cardiologica
|March 22, 2022
PubMed

Insights

The Select Secure® lead is safe and effective for paediatric patients needing cardiac devices. This transvenous pacing system demonstrated stable electrical performance and feasibility in children over midterm follow-up.

Area of Science:

  • Cardiology
  • Biomedical Engineering
  • Paediatric Electrophysiology

Background:

  • Cardiac device implantation in paediatric populations presents unique challenges.
  • Evaluating novel pacing lead systems for safety and efficacy in children is crucial.

Purpose of the Study:

  • To assess the safety, feasibility, and efficacy of the Select Secure® lead in a paediatric cohort.
  • To document implantation techniques, complications, and acute/midterm performance.

Main Methods:

  • A retrospective analysis of 42 paediatric patients (78 leads) implanted between 2014-2020.
  • Evaluation of patient demographics, implantation details, sensing/pacing characteristics, and impedances.
  • Monitoring for complications during acute and midterm follow-up (median 47 months).

Main Results:

  • All 78 Select Secure® leads were successfully implanted, with ventricular leads anchored to the right ventricular midseptum.
  • Implantation procedures were uneventful, with no major complications reported.
  • Electrical parameters (sensing, pacing thresholds, impedance) remained stable throughout the follow-up period.

Conclusions:

  • The Select Secure® transvenous pacing system is safe and feasible for paediatric cardiac device implantation.
  • The system exhibits very stable electrical performance in the acute and midterm phases.
  • This study supports the use of Select Secure® leads in paediatric patients.

Related Concept Videos