Adaptive dual-energy algorithm based on pre-calibrated weighting factors for chest radiography

Ivan Romadanov1, Ruwan Abeywardhana1, Mike Sattarivand1,2,3

  • 1Department of Medical Physics, Nova Scotia Health Authority, Halifax, NS, Canada.

Summary

This study introduces a new dual-energy imaging algorithm that adapts to spatial variations in tissue composition. The method uses pre-calibrated weighting factors to improve material selection and noise suppression in chest radiography. Calibration phantoms were used to determine optimal weighting factors for soft-tissue and bone separation. The algorithm was tested on Rando phantoms in different orientations and sizes. Results showed that the adaptive algorithm improved image quality compared to conventional methods. The adaptive anti-correlated noise reduction (aACNR) further reduced noise but slightly decreased contrast. The findings suggest that this approach can enhance diagnostic accuracy in clinical settings by providing clearer material differentiation.