3D CAIPIRINHA SPACE versus standard 2D TSE for routine knee MRI: a large-scale interchangeability study
Pieter Van Dyck1,2, Céline Smekens3, Ella Roelant4
1Department of Radiology, Antwerp University Hospital (UZA), Drie Eikenstraat 655, B-2650, Edegem, Belgium. pieter.van.dyck@uza.be.
Objective:
To perform a large-scale interchangeability study comparing 3D controlled aliasing in parallel imaging results in higher acceleration (CAIPIRINHA) sampling perfection with application optimized contrast using different flip angle evolutions (SPACE) TSE with standard 2D TSE for knee MRI.
Methods:
In this prospective study, 250 patients underwent 3 T knee MRI, including a multicontrast 3D CAIPIRINHA SPACE TSE (9:26 min) and a standard 2D TSE protocol (12:14 min). Thirty-three (13%) patients had previous anterior cruciate ligament and/or meniscus surgery. Two radiologists assessed MRIs for image quality and identified pathologies of menisci, ligaments, and cartilage by using a 4-point Likert scale according to the level of diagnostic confidence. Interchangeability of the protocols was tested under the same-reader scenario using a bootstrap percentile confidence interval. Interreader reliability and intermethod concordance were also evaluated.
Results:
Despite higher image quality and diagnostic confidence for standard 2D TSE compared to 3D CAIPIRINHA SPACE TSE, the protocols were found interchangeable for diagnosing knee abnormalities, except for patellar (6.8% difference; 95% CI: 4.0, 9.6) and trochlear (3.6% difference; 95% CI: 0.8, 6.6) cartilage defects. The interreader reliability was substantial to almost perfect for 2D and 3D MRI (range κ, 0.785-1 and κ, 0.725-0.964, respectively). Intermethod concordance was almost perfect for all diagnoses (range κ, 0.817-0.986).
Conclusion:
Multicontrast 3D CAIPIRINHA SPACE TSE and standard 2D TSE protocols perform interchangeably for diagnosing knee abnormalities, except for patellofemoral cartilage defects. Despite the radiologist's preference for 2D TSE imaging, a pursuit towards time-saving 3D TSE knee MRI is justified for routine practice.
Key Points:
• Multicontrast 3D CAIPIRINHA SPACE and standard 2D TSE protocols perform interchangeably for diagnosing knee abnormalities, except for patellofemoral cartilage defects. • Radiologists are more confident in diagnosing knee abnormalities on 2D TSE than on 3D CAIPIRINHA SPACE TSE MRI. • Despite the radiologist's preference for 2D TSE, a pursuit towards accelerated 3D TSE knee MRI is justified for routine practice.
More Related Videos
06:59Improved Registration of 3D CT Angiography with X-ray Fluoroscopy for Image Fusion During Transcatheter Aortic Valve Implantation
Published on: June 3, 2018
09:41A Pipeline for 3D Multimodality Image Integration and Computer-assisted Planning in Epilepsy Surgery
Published on: May 20, 2016
Related Concept Videos
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies IV: Magnetic Resonance Imaging
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...