Risk factors for hemolysis with centrifugal pumps in pediatric extracorporeal membrane oxygenation: Is pump
Jian H Chu1,2, Srivats Sarathy3, Sonali Ramesh4
1University of Iowa Carver College of Medicine, Iowa City, IA 52242, USA.
Introduction:
Hemolysis during pediatric extracorporeal membrane oxygenation (ECMO) is associated with increased risk for renal failure and mortality.
Objectives:
We aim to describe risk factors for hemolysis in pediatric ECMO supported by centrifugal pumps.
Methods:
We conducted an analysis of retrospective data collected at an academic children's hospital from January 2017 to December 2019.
Measurements And Results:
Plasma-free hemoglobin (PFH) levels were measured daily, and hemolysis was defined as PFH>50 mg/dL. Of 46 ECMO runs over 528 ECMO days, hemolysis occurred in 23 (58%) patients over a total of 40 (8%) ECMO days. In multivariable logistic regression models, VA-ECMO (aOR=4.69, 95% CI: 1.01-21.83) and higher hemoglobin (aOR = 1.38, 95% CI: 1.06-1.81) were independently associated with hemolysis. There were also non-significant trends toward increased risk for hemolysis with higher rotational pump speed (aOR=2.39, 95% CI: 0.75-7.65), higher packed red blood cell transfusions (aOR=1.15, 95% CI: 0.99-1.34), and higher cryoprecipitate transfusions (aOR=2.01, 95% CI: 0.83-4.86). Isolated pump exchanges that were performed in 12 patients with hemolysis led to significant decreases in PFH levels within 24 h (89 vs 11 mg/dL, p<0.01).
Conclusions:
Hemolysis is common in pediatric ECMO using centrifugal pumps. Avoidance of high pump speeds and conservative administration of blood products may help to mitigate the risk for hemolysis. Furthermore, pump exchange may be an effective first-line treatment for hemolysis.
More Related Videos
07:30In Vitro Model of Physiological and Pathological Blood Flow with Application to Investigations of Vascular Cell Remodeling
Published on: November 3, 2015
09:47Author Spotlight: Utilizing Next-Generation Polymerized Human Hemoglobin for Improved Donor Lung Evaluation and Preservation in Rats
Published on: June 14, 2024
Related Concept Videos
Hemodialysis II: Procedure and Complications
Hemodialysis I: Introduction
Continuous Renal Replacement Therapy
Peritoneal Dialysis II: Peritoneal Dialysis Systems and Complications