The use of pseudo-MRM for a sensitive and selective detection and quantification of polycyclic aromatic compounds by
Mathieu Galmiche1,2, Anaïs Rodrigues1, Estelle Motsch3
1Institut de Chimie et Procédés pour l'Énergie, l'Environnement et la Santé (ICPEES) - Physico-Chimie de l'Atmosphère, Université de Strasbourg, Strasbourg, France.
Rationale:
Multiple Reaction Monitoring (MRM) is a sensitive and selective detection mode for target trace-level analysis. However, it requires the fragmentation of labile bonds which are not present in molecules such as Polycyclic Aromatic Hydrocarbons (PAHs) and their heterocyclic derivatives (PANHs, PASHs).
Methods:
We present the application of an alternative tandem mass spectrometry (MS/MS) mode called "pseudo-MRM" for the GCMS/MS analysis of Polycyclic Aromatic Compounds (PACs). This mode is based on the monitoring of transitions with no mass loss between the precursor and the product ion. Pseudo-MRM peak areas were compared with those of classic MRM on three different mass spectrometers: two triple quadrupoles and an ion trap.
Results:
For all non-polar PACs studied here (PAHs, PANHs and PASHs), the pseudo-MRM transition was always the most intense. The classic MRM transitions exhibited peak areas 2 to 5 times lower. On the contrary, for the functionalized PACs (oxygenated and nitrated PAHs), classic MRM was favored over pseudo-MRM. These observations were confirmed on two triple quadrupoles (QqQs), and the real-world applicability of pseudo-MRM on QqQs was validated by the successful analysis of Diesel PM. However, a comparison with an ion trap showed that pseudo-MRM was never favored on that instrument, which caused fragmentation of non-polar PACs in MS/MS.
Conclusions:
The results of this study show an important gain in sensitivity when using pseudo-MRM instead of MRM for non-polar PACs on QqQ instruments. The selectivity of MRM is preserved in pseudo-MRM by applying non-zero collision energies to which only these non-polar PACs are resistant, not the isobaric interferences. No interference issue was observed when analyzing Diesel PM, a complex matrix, with our pseudo-MRM method. Therefore, we advise for a broader use of this MS/MS mode for trace-level determination of non-polar PAHs.
More Related Videos
09:26Species Determination and Quantitation in Mixtures Using MRM Mass Spectrometry of Peptides Applied to Meat Authentication
Published on: September 20, 2016
11:09An HS-MRM Assay for the Quantification of Host-cell Proteins in Protein Biopharmaceuticals by Liquid Chromatography Ion Mobility QTOF Mass Spectrometry
Published on: April 17, 2018
Related Concept Videos
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...
High-Resolution Mass Spectrometry (HRMS)
Peptide Identification Using Tandem Mass Spectrometry
This technique helps gather information regarding the protein from which the peptide was obtained and to study the peptides’ amino acid sequence. Identifying peptides from a complex mixture is an important component of the growing field of...
Mass Spectrometry: Aromatic Compound Fragmentation
Gas Chromatography: Types of Detectors-II
¹³C NMR: Distortionless Enhancement by Polarization Transfer (DEPT)