Jove
Visualize
Contact Us
JoVE
x logofacebook logolinkedin logoyoutube logo
ABOUT JoVE
OverviewLeadershipBlogJoVE Help Center
AUTHORS
Publishing ProcessEditorial BoardScope & PoliciesPeer ReviewFAQSubmit
LIBRARIANS
TestimonialsSubscriptionsAccessResourcesLibrary Advisory BoardFAQ
RESEARCH
JoVE JournalMethods CollectionsJoVE Encyclopedia of ExperimentsArchive
EDUCATION
JoVE CoreJoVE BusinessJoVE Science EducationJoVE Lab ManualFaculty Resource CenterFaculty Site
Terms & Conditions of Use
Privacy Policy
Policies

Related Concept Videos

Tandem Mass Spectrometry01:21

Tandem Mass Spectrometry

1.3K
Tandem mass spectrometry is a technique that uses multiple mass analyzers in series to obtain a higher selectivity and signal-to-noise ratio for the analyte. Instruments with multiple analyzers separated by an interaction cell enable secondary fragmentation and selected study of the fragment ions.
Secondary fragmentations occur in the interaction cell and can be induced by various factors. Fragmentation induced by collision with inert gases, such as N2, Ar, He, etc., is called collision-induced...
1.3K
Mass Spectrometry: Complex Analysis01:21

Mass Spectrometry: Complex Analysis

920
Mass spectrometry is an important technique for the identification of pure compounds. However, it has some limitations for the analysis of complex mixtures, often due to excessive fragmentation making the spectrum too complicated to decipher. Mass spectrometry can be combined with suitable separation methods in sequence, forming hyphenated methods, which are useful in the analysis of complex mixtures.
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...
920
Gas Chromatography–Mass Spectrometry (GC–MS)01:14

Gas Chromatography–Mass Spectrometry (GC–MS)

4.9K
Gas chromatography–mass spectrometry (GC–MS) is the combination of analytical techniques of gas chromatography and mass spectrometry in a single instrument for analyzing a mixture of compounds. The gas chromatograph separates the compounds in the mixture, and the mass spectrometer analyzes each compound separately to determine the molecular masses and molecular structures.
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall....
4.9K
Mass Analyzers: Common Types01:19

Mass Analyzers: Common Types

727
The quadrupole mass analyzer consists of four cylindrical metal rods arranged in a diamond carrying a DC voltage and a radio-frequency AC voltage. The motion of ions through the quadrupole depends on the field strength, causing only ions of a certain m/z to resonate successfully and strike the detector at a given field strength. Though the transmission rate for these analyzers is high, the exact elemental composition of the sample is not determined because of low resolution; however, they are...
727
Peptide Identification Using Tandem Mass Spectrometry01:33

Peptide Identification Using Tandem Mass Spectrometry

7.0K
Tandem mass spectrometry, also known as MS/MS or MS2, is an analytical technique that employs two mass analyzers. Essentially it is a series of mass spectrometers that helps isolate a particular biomolecule and then helps study its chemical properties.
This technique helps gather information regarding the protein from which the peptide was obtained and to study the peptides’ amino acid sequence. Identifying peptides from a complex mixture is an important component of the growing field of...
7.0K
Mass Spectrometers01:16

Mass Spectrometers

6.1K
This lesson details the instrumentation of a mass spectrometer—a physical instrument to perform mass spectrometry on analyte molecules and record the characteristic mass spectra. This is achieved via three chief functions:
6.1K

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Detection and Characterization of Plasma-Generated Stannane: Influence of Surface Composition on Species Formation.

Rapid communications in mass spectrometry : RCM·2025
Same author

Signatures of Charged Droplets from ESI: A Statistical Analysis of Non-summed Mass Spectra Compared to APCI.

Journal of the American Society for Mass Spectrometry·2025
Same author

Investigation of Surface-Induced Dissociation Processes via Molecular Dynamics Simulations of Wall Collisions of Large Droplets Produced by Electrospray Ionization.

Journal of the American Society for Mass Spectrometry·2025
Same author

New particle formation from isoprene under upper-tropospheric conditions.

Nature·2024
Same author

Simulation of Space Charge Effects in Fourier Transform Quadrupole Ion Traps (FT-QITs).

Journal of the American Society for Mass Spectrometry·2024
Same author

Dynamics of the Aspiration of Charged Droplets into a LC-ESI-MS System.

Journal of the American Society for Mass Spectrometry·2024

Related Experiment Video

Updated: Sep 27, 2025

Using a Cyclic Ion Mobility Spectrometer for Tandem Ion Mobility Experiments
08:40

Using a Cyclic Ion Mobility Spectrometer for Tandem Ion Mobility Experiments

Published on: January 20, 2022

4.4K

Parallel Operation of Electron Ionization and Chemical Ionization for GC-MS Using a Single TOF Mass Analyzer.

Steffen Bräkling1,2, Kai Kroll2, Carsten Stoermer1

  • 1TOFWERK, Thun 3645, Switzerland.

Analytical Chemistry
|April 7, 2022
PubMed
Summary

This study introduces a novel gas chromatography-mass spectrometry (GC-MS) system that combines electron ionization (EI) and chemical ionization (CI) for enhanced compound identification. The dual-ionization approach provides complementary spectral data from a single run.

More Related Videos

High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
10:17

High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry

Published on: April 23, 2019

9.8K
Measurement of Ultrafast Vibrational Coherences in Polyatomic Radical Cations with Strong-Field Adiabatic Ionization
08:22

Measurement of Ultrafast Vibrational Coherences in Polyatomic Radical Cations with Strong-Field Adiabatic Ionization

Published on: August 6, 2018

7.0K

Related Experiment Videos

Last Updated: Sep 27, 2025

Using a Cyclic Ion Mobility Spectrometer for Tandem Ion Mobility Experiments
08:40

Using a Cyclic Ion Mobility Spectrometer for Tandem Ion Mobility Experiments

Published on: January 20, 2022

4.4K
High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
10:17

High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry

Published on: April 23, 2019

9.8K
Measurement of Ultrafast Vibrational Coherences in Polyatomic Radical Cations with Strong-Field Adiabatic Ionization
08:22

Measurement of Ultrafast Vibrational Coherences in Polyatomic Radical Cations with Strong-Field Adiabatic Ionization

Published on: August 6, 2018

7.0K

Area of Science:

  • Analytical Chemistry
  • Mass Spectrometry
  • Chromatography

Background:

  • Traditional GC-MS often relies on a single ionization technique (EI or CI), limiting the scope of spectral information.
  • Obtaining complementary EI and CI data typically requires separate analyses, increasing time and resource consumption.

Purpose of the Study:

  • To develop and validate a novel GC-MS system capable of simultaneously acquiring both EI and CI mass spectra for a single chromatographic peak.
  • To demonstrate the enhanced analytical capabilities for compound identification using this dual-ionization approach.

Main Methods:

  • A novel GC-MS instrument was designed with a parallel split to electron ionization (EI) and chemical ionization (CI) sources.
  • A time-of-flight (TOF) mass analyzer was coupled with fast ion optical switching and synchronized data acquisition.
  • The system was tested using standard GC mixtures and a complex perfume sample.

Main Results:

  • The system successfully acquired complementary EI and CI mass spectral information within a single chromatographic run.
  • High sensitivity was achieved with instrument detection limits of <40 fg (EI) and <2 pg (CI).
  • Automatic spectral alignment of EI and CI data was demonstrated, simplifying analysis.

Conclusions:

  • The developed dual-EI/CI GC-MS system offers a significant advancement in analytical chemistry, providing richer data for compound identification.
  • This technique enhances chromatographic fidelity and analytical value, particularly for complex samples.
  • The system's performance validates its utility in routine analysis and discovery.