Related Experiment Video
Updated: Sep 27, 2025

Transcutaneous Microcirculatory Imaging in Preterm Neonates
Published on: December 31, 2015
Evaluation of a new transcutaneous bilirubinometer in newborn infants
Mikael Norman1,2, Hüseyin Aytug3, Hasan Basri Celebi4
1Division of Pediatrics, Department of Clinical Science, Intervention and Technology, Karolinska Institutet, Novum, Blickagången 6A, 141 57, Stockholm, Sweden. mikael.norman@ki.se.
Abstract:
To avoid brain damage in newborn infants, effective tools for prevention of excessive neonatal hyperbilirubinemia are needed. The objective of this study was to evaluate a new transcutaneous bilirubinometer (JAISY). For this purpose, 930 bilirubin measurements were performed in 141 newborn infants born near-term or at term (gestational age 35-41 weeks; postnatal age 1-6 days; 71 boys; including 29 infants with darker skin) and compared to those of a previously validated instrument (JM105). In each infant, the mean of three repeated measurements in the forehead was calculated for each instrument, followed by a similar measurement on the chest. The bilirubin values varied between 0 and 320 µmol/l (0-18.8 mg/dl). There was a high degree of agreement with significant correlations between bilirubin values measured with the two devices on the forehead (Pearson's r = 0.94, p < 0.001) and the chest (r = 0.94, p < 0.001). The correlations remained after stratifying the data by gestational age, postnatal age and skin color. The coefficient of variation for repeated bilirubin measurements was 8.8% for JAISY and 8.0% for JM105 (p = 0.79). In conclusion, JAISY provides accurate and reproducible information on low to moderately high bilirubin levels in newborn infants born near-term or at term.