True-positive reflex threshold value for HCV antibody screening test
Ainulkhir Hussin1, Rusmah Yusof1, Mohd Yusof Nor Rahim1
1Department of Pathology, Queen Elizabeth Hospital, Ministry of Health, Kota Kinabalu, Sabah, Malaysia.
Introduction:
The use of a signal-to-cut-off ratio has been recommended by the Centre for Disease Control and Prevention to determine the need for further validation using a supplemental test. In this study, we aimed to determine the optimal true-positive signal-to-cut-off ratio for the ABBOTT ARCHITECT i2000SR immunoassay (Abbott Laboratories, Illinois, USA), using the Serodia® HCV particle agglutination (HCV-PA) assay (Fujirebio Inc, Tokyo, Japan) as the reference test for anti-HCV screening.
Methodology:
We analysed a total of 13,240 specimens using the ARCHITECT i2000SR immunoassay and subsequently subjected all the reactive specimens with a signal-to-cut-off ratio ≥ 1.00 (n = 267) to the Serodia® HCV-PA reference assay. Receiver operating characteristic (ROC) curve analysis was carried out and performance characteristics for each signal-to-cut-off ratio were determined. The selected signal-to-cut-off ratio value was then assessed using a line immunoassay (LIA) test.
Results:
ROC curve analysis determined that the optimal signal-to-cut-off ratio was 5.05, which gave the highest Youden's Index (J) value of 0.89, with a sensitivity of 93.1% (88.9-97.2), a specificity of 96.0% (92.4-99.4), a positive predictive value of 96.4% (93.3-99.5), and a negative predictive value of 92.2% (87.5-96.8). Validation of the optimal S/Co value using the LIA test yielded an accuracy of 91.8%, with sensitivity and specificity values of 92.0% and 91.7%, respectively.
Conclusions:
The optimal signal-to-cut-off ratio value for the ARCHITECT i2000SR immunoassay, which was determined using HCV-PA assay as the reference test and validated using a HCV-LIA assay, showed high sensitivity and specificity, and may be used in routine anti-HCV screening.