Related Experiment Video
Updated: Sep 27, 2025

Proton Therapy Delivery and Its Clinical Application in Select Solid Tumor Malignancies
Published on: February 6, 2019
Development and dosimetric verification of 3D customized bolus in head and neck radiotherapy
Nichakan Chatchumnan1, Sakda Kingkaew1, Chuanchom Aumnate2
1Division of Radiation Oncology, Department of Radiology, King Chulalongkorn Memorial Hospital, Bangkok 10330, Thailand.
Abstract:
The commercial flat bolus cannot form perfect contact with the irregular surface of the patient's skin, resulting in an air gap. The purpose of this study was to evaluate the feasibility of using a 3D customized bolus from silicone rubber. The silicone rubber boluses were studied in basic characteristics. The 3D customized bolus was fabricated at the nose, cheek and neck regions. The point dose and planar dose differences were evaluated by comparing with virtual bolus. The hardness, thickness, density, Hounsfield unit (HU) and dose attenuation of the customized bolus were quite similar to a commercial bolus. When a 3D customized bolus was placed on the RANDO phantom, it can significantly increase buildup region doses and perfectly fit against the irregular surface shape. The average point dose differences of 3D customized bolus were -1.1%, while the commercial bolus plans showed -1.7%. The average gamma results for planar dose differences comparison of 3D customized bolus were 93.9%, while the commercial bolus plans were reduced to 91.9%. Overall, A silicone rubber bolus produced the feasible dosimetric properties and could save cost compared to a commercial bolus. The 3D printed customized bolus is a good buildup material and could potentially replace and improve treatment efficiency.

