Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
Small static radiosurgery field dosimetry with small volume ionization chambers
Miguel López-Sánchez1, María Pérez-Fernández2, Eduardo Pardo3
1Hospital Universitario Lucus Augusti, R. Dr. Ulises Romero, 1, E-27003, Lugo, Spain; Radiation Physics Laboratory, AII, University of Santiago de Compostela, E. San Lourenzo S/N, E-15705, Santiago de Compostela, Spain.
Purpose:
To evaluate the response of the four smallest active volume thimble type ionization chambers commercially available (IBA-dosimetry RAZOR Nano Chamber, Standard Imaging Exradin A16, IBA-dosimetry CC01 and PTW T31022) when measuring SRS cone collimated Flattening Filter Free (FFF) fields.
Methods:
We employed Monte Carlo simulation for calculating correction factors as defined in IAEA TRS-483. Monte Carlo simulation beam model and ion chamber geometry definitions were supported by an extensive set of measurements. Type A and B uncertainty components were evaluated.
Results:
Commissioning of Monte Carlo 6 MV and 10 MV FFF beam models yielded relative differences between measured and simulated dose distributions lower than 1.5%. Monte Carlo simulated output factors for 5 mm SRS field agree with experimental values within 1% local relative difference for all chambers. Smallest active volume ion chamber (IBA-dosimetry RAZOR Nano Chamber) exhibits smallest correction, being compatible with unity. Correction factor combined uncertainties range between 0.7% and 0.9%. Smallest uncertainties were recorded for smallest and largest active volume ion chambers, although the latter exhibited largest correction factor. Highest contribution to combined uncertainty was type B component associated with beam model initial electron spatial Full Width Half Maximum (FWHM) uncertainty.
Conclusions:
Among the investigated chambers, the IBA RAZOR Nano Chamber was found to be an excellent choice for narrow beam output factor measurement since it requires minimum correction (in line with IAEA TRS-483 recommendations). This is caused by its tiny size and tissue equivalence materials which produce minimum volume averaging and fluence perturbation.